Message ID: 601     Entry time: Monday, February 17, 2014, 09:32
Author: David Wang 
Category: Conditioning Station 
Type: Repair 
Module: TM2 
Target/Number:  
Subject: helium pressure test on TM2 tube heater +(D) water line. 

I did the helium pressure test on Tube heater+ line this morning. The vacuum on CS and TM2 are very good. The base lake rate on leak detector was 1.7xE-7 atm.cc/sec. Base pressure was 0.0xE-4torr.  I applied helium to the "D"line .The test started from 30psi up to 80psi on both side "D" line.The base leak rate slightly changed to 1.9xE-7 atm.cc/sec during the test,but not base pressure. At this base leak rate level, No obvious leak had been found . See attachment pictures. The test will be repeated later when the base leak rate could be pumped into E -8 or E-9(the best) range.            

Attachment 1: 002.JPG  2.677 MB  | Show | Hide all
Attachment 2: 003.JPG  2.261 MB  | Show | Hide all
Attachment 3: 004.JPG  2.254 MB  | Show | Hide all
Attachment 4: 012.JPG  3.421 MB  | Show | Hide all
Attachment 5: 016.JPG  2.612 MB  | Show | Hide all
Attachment 6: 018.JPG  2.339 MB  | Hide | Hide all