Message ID: 601
Entry time: Monday, February 17, 2014, 09:32
|
Author: |
David Wang |
Category: |
Conditioning Station |
Type: |
Repair |
Module: |
TM2 |
Target/Number: |
|
Subject: |
helium pressure test on TM2 tube heater +(D) water line. |
|
|
I did the helium pressure test on Tube heater+ line this morning. The vacuum on CS and TM2 are very good. The base lake rate on leak detector was 1.7xE-7 atm.cc/sec. Base pressure was 0.0xE-4torr. I applied helium to the "D"line .The test started from 30psi up to 80psi on both side "D" line.The base leak rate slightly changed to 1.9xE-7 atm.cc/sec during the test,but not base pressure. At this base leak rate level, No obvious leak had been found . See attachment pictures. The test will be repeated later when the base leak rate could be pumped into E -8 or E-9(the best) range. |
|
|
|
|
|
|
|