Thursday, February 13, 2014, 13:30, David Wang, South Hot-Cell, Repair, TM2, , pressure test on TM2 tube heater +(D) water line.
|
Thursday, February 13, 2014, 14:15, David Wang, South Hot-Cell, Repair, TM2, , pressure test on TM2 tube heater +(D) water line.
|
|
Message ID: 599
Entry time: Thursday, February 13, 2014, 14:15
In reply to: 598
|
Author: |
David Wang |
Category: |
South Hot-Cell |
Type: |
Repair |
Module: |
TM2 |
Target/Number: |
|
Subject: |
pressure test on TM2 tube heater +(D) water line. |
|
|
David Wang wrote: |
Chad and I did the pressure test on TM2tube heater+(D) water line at SHC with shop air.We tested from 30psi to 80psi on both side of D line. The bubbles were found at 50-55psi on blank off seal in SHC.
|
I would like to correct this E-LOG. There is no real leaking bubble had been found during the whole test according to the information supplied by Chad Fisher during this afternoon meeting. I would like to attach some pictures took from TH during this test. |
|
|
|