Message ID: 597
Entry time: Thursday, February 13, 2014, 09:24
|
Author: |
David Wang |
Category: |
South Hot-Cell |
Type: |
Repair |
Module: |
TM2 |
Target/Number: |
|
Subject: |
Leak check TM2 tube heater (+) water line with rubber insert blocked WB in SHC |
|
|
Chad and I did the Leak checked TM2 tube heater (+) water line with rubber insert plugged water block on " D" in SHC. The Leak detector was very easy to pump down to the bottom leak rate and pressure. Chad checked the helium flow in SHC, and sprayed 10 second helium on the block. There is no response had been found during the test. The test was repeated again and no leak has been found. See attachment pictures. |
|
|
|
|