Message ID: 591
Entry time: Tuesday, February 04, 2014, 08:04
|
Author: |
David Wang |
Category: |
South Hot-Cell |
Type: |
Repair |
Module: |
TM2 |
Target/Number: |
|
Subject: |
Leak check TM2 tube heater (+) water line. |
|
|
Chad and I leak checked TM2 tube heater(+) water line yesterday afternoon. The water line was difficult to pump down on base leak rate, but LD base pressure was fine . The base pressure on leak detector was pumped to 0.0XE-4 torr quickly. The base leak rate only could be pumped to 4.5 xE-8 atm.cc/sec after 1 hour waiting time. All vacuum connections to leak detector in the target hall were leak checked and they were good. Chad checked helium flow in SHC. He sprayed helium twice on tube heater(+) water block soldering joints. I did not see any response on leak detector in TH.The base Leak rate and base pressure did not move up during the test. See attachment pictures. |
|
|
|