TM2 no source tray 55kV 1 hour |
|
|
Today 55kV was applied to TM2 and held for over 1 hour with the EINZEL at 27.5kV. There were 2 sparks and the current was unstable at this voltage but this test should be considered a success. Once again while bringing the voltage up the canbus continued to be a problem. After successfully passing the test I continued to raise the BIAS and EINZEL. A spark at 57kV tripped a controller for the vacuum and the canbus. As stated before I believe the module will reach 60kV and that there is an interaction between the EINZEL and the BIAS. I am not certain if that interaction is internal or external to the module. Attached is a screen shot and the dumped data. |
|
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:DIC1:RDCOND [Mohm] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
12/06/2013 10:34:30.138858 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:31.138402 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:32.138289 0.000000093 0.000001010 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:33.138127 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:34.138033 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:35.137874 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:36.137749 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:37.137657 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:38.137550 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:39.138452 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:40.138322 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:41.138150 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:42.139015 0.000000094 0.000000996 0.000000768 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:43.138906 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:44.138797 0.000000093 0.000000996 0.000000765 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:45.138654 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:46.144533 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:47.144548 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:48.144363 0.000000093 0.000000998 0.000000768 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:49.144158 0.000000093 0.000000996 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:50.144148 0.000000093 0.000000996 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:51.144894 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:52.144785 0.000000093 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:53.144636 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:54.144483 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:55.157531 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:56.158254 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:57.158119 0.000000094 0.000001032 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:58.157969 0.000000094 0.000001010 0.000000768 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:59.158877 0.000000093 0.000000998 0.000000768 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:00.159762 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:01.159615 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:03.306946 0.000000093 0.000000996 0.000000768 16.9 BadVal BadVal BadVal BadVal
12/06/2013 10:35:04.307252 0.000000094 0.000000998 0.000000768 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:05.308070 0.000000094 0.000000998 0.000000768 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:06.308956 0.000000094 0.000000994 0.000000770 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:07.308826 0.000000094 0.000000994 0.000000770 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:08.308694 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:09.308573 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:10.309442 0.000000094 0.000001010 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:11.309317 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:12.309197 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:13.309123 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:14.308964 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:15.309987 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:16.309785 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:17.309567 0.000000094 0.000000996 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:18.422297 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:19.422322 0.000000094 0.000000994 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:20.422292 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:21.422265 0.000000094 0.000001010 0.000000772 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:22.422985 0.000000094 0.000001010 0.000000772 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:23.532598 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:24.532696 0.000000094 0.000000996 0.000000772 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:25.700311 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:26.754388 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:27.754303 0.000000094 0.000000996 0.000000768 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:28.754164 0.000000094 0.000000996 0.000000768 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:29.808485 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:30.808903 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:31.892662 0.000000093 0.000000996 0.000000772 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:32.892649 0.000000094 0.000000994 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:33.892524 0.000000094 0.000000994 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:34.892399 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:36.59457 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:37.59452 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:38.58995 0.000000094 0.000000998 0.000000768 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:35:39.58868 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:40.58702 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:41.59611 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:42.59451 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:43.59465 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:44.69481 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 8.54 0.01
12/06/2013 10:35:45.70113 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 8.54 0.01
12/06/2013 10:35:46.69968 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:47.69843 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:48.69730 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:49.69609 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:50.69501 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:51.70349 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:52.70327 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:53.71064 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:54.71157 0.000000094 0.000000996 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:55.70854 0.000000093 0.000000998 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:56.70728 0.000000093 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:57.70696 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:58.71465 0.000000094 0.000000998 0.000000772 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:59.71304 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:36:00.72214 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:36:01.72058 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:02.72037 0.000000094 0.000000996 0.000000765 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:03.71845 0.000000093 0.000000996 0.000000768 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:36:04.71718 0.000000093 0.000000996 0.000000770 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:36:05.71547 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:06.72545 0.000000094 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:07.72467 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:08.73208 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:09.73087 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:10.73946 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:11.74864 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:12.74666 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:13.75645 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:14.75641 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:15.75369 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:16.136319 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:17.136267 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:18.230037 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:19.229817 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:20.229681 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:21.229580 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:22.229431 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:23.229311 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:24.229201 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:25.229010 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:26.228930 0.000000094 0.000000994 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:27.273591 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:28.273699 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:29.274782 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:30.274160 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:31.274279 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:32.274186 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:33.276910 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:34.276923 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:35.277751 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:36.277627 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:37.277496 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:38.277402 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:39.277265 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:40.277272 0.000000093 0.000001010 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:41.278116 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:42.277876 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:43.277804 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:44.277667 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:45.277528 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:46.279324 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:47.279479 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:48.280171 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:49.280032 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:50.279918 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:51.280727 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:52.280665 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:53.280477 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:54.280361 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:55.280370 0.000000094 0.000000994 0.000000770 16.7 27.46 0.06 9.76 0.01
12/06/2013 10:36:56.281267 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:57.281021 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:58.280911 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:59.280785 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:00.280647 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:01.281537 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:02.281408 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:03.281254 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:04.281106 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:05.281008 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:06.281889 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:07.281764 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:08.281734 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:09.281500 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:10.281385 0.000000094 0.000000994 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:11.282277 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.76 0.01
12/06/2013 10:37:12.282243 0.000000093 0.000000998 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:13.282030 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:14.281905 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:15.281781 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:16.282607 0.000000093 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:17.282622 0.000000093 0.000000998 0.000000768 16.9 6.86 1.03 9.15 0.01
12/06/2013 10:37:18.282434 0.000000094 0.000000998 0.000000770 16.9 6.86 1.03 1.22 0.01
12/06/2013 10:37:19.282257 0.000000094 0.000000998 0.000000770 16.9 3.43 1.03 1.22 0.01
12/06/2013 10:37:20.282499 0.000000094 0.000000998 0.000000770 16.9 4.57 1.03 1.83 0.01
12/06/2013 10:37:21.283005 0.000000094 0.000000998 0.000000770 16.9 4.57 1.03 493.17 0.01
12/06/2013 10:37:22.283870 0.000000093 0.000000998 0.000000770 16.9 4.57 1.03 993.05 0.01
12/06/2013 10:37:23.283923 0.000000094 0.000000998 0.000000770 16.9 6.86 0.91 1494.16 0.01
12/06/2013 10:37:24.283648 0.000000093 0.000000996 0.000000770 16.9 3.43 1.15 1494.16 0.01
12/06/2013 10:37:25.283492 0.000000094 0.000000998 0.000000770 16.9 13.73 1.15 1994.04 0.28
12/06/2013 10:37:26.283509 0.000000094 0.000000998 0.000000770 16.9 13.73 1.15 1994.04 0.08
12/06/2013 10:37:27.283262 0.000000093 0.000000998 0.000000770 16.8 5.72 0.85 2494.54 0.10
12/06/2013 10:37:28.283071 0.000000093 0.000000998 0.000000770 16.8 2.28 0.97 2995.04 0.10
12/06/2013 10:37:29.282990 0.000000094 0.000000998 0.000000770 16.8 2.28 0.97 3495.53 0.10
12/06/2013 10:37:30.282877 0.000000093 0.000000998 0.000000770 16.8 10.29 0.97 3994.81 0.15
12/06/2013 10:37:31.283728 0.000000094 0.000000996 0.000000768 16.8 8.01 0.91 3994.81 0.15
12/06/2013 10:37:32.283568 0.000000094 0.000000998 0.000000768 16.8 4.57 1.15 3996.03 0.15
12/06/2013 10:37:33.283508 0.000000093 0.000000998 0.000000770 16.8 6.86 1.15 3996.03 0.15
12/06/2013 10:37:34.283108 0.000000094 0.000000998 0.000000768 16.8 4.57 1.03 3996.03 0.15
12/06/2013 10:37:35.283283 0.000000094 0.000000998 0.000000768 16.8 5.72 1.03 3996.03 0.15
12/06/2013 10:37:36.284116 0.000000093 0.000000994 0.000000768 16.8 5.72 1.09 3996.03 0.15
12/06/2013 10:37:37.283983 0.000000094 0.000000998 0.000000770 16.8 5.72 1.09 3996.64 0.15
12/06/2013 10:37:38.283864 0.000000094 0.000000998 0.000000770 16.7 3.43 1.09 3996.03 0.15
12/06/2013 10:37:39.472917 0.000000094 0.000000998 0.000000765 16.7 3.43 1.09 3996.64 0.15
12/06/2013 10:37:40.472581 0.000000093 0.000000996 0.000000768 16.7 2.28 1.15 3996.64 0.15
12/06/2013 10:37:41.472472 0.000000094 0.000000998 0.000000770 16.7 2.28 1.15 3996.64 0.15
12/06/2013 10:37:42.472333 0.000000094 0.000000998 0.000000768 16.8 8.01 1.03 3996.64 0.15
12/06/2013 10:37:43.472212 0.000000094 0.000000998 0.000000770 16.8 8.01 1.03 3996.64 0.15
12/06/2013 10:37:44.738171 0.000000094 0.000000998 0.000000770 16.8 9.15 1.03 3996.64 0.15
12/06/2013 10:37:45.737929 0.000000093 0.000000998 0.000000770 16.8 9.15 1.03 3996.03 0.15
12/06/2013 10:37:46.737758 0.000000093 0.000000996 0.000000768 16.8 10.29 1.03 3996.64 0.15
12/06/2013 10:37:47.737679 0.000000094 0.000000996 0.000000770 16.8 8.01 0.91 3996.64 0.15
12/06/2013 10:37:48.737550 0.000000094 0.000000998 0.000000768 16.8 5.72 1.15 3996.64 0.15
12/06/2013 10:37:49.737448 0.000000094 0.000000998 0.000000768 16.8 5.72 1.15 4276.79 0.15
12/06/2013 10:37:50.737254 0.000000094 0.000000998 0.000000770 16.8 3.43 0.97 4997.02 0.15
12/06/2013 10:37:51.738371 0.000000093 0.000000996 0.000000770 16.8 12.58 0.91 5997.40 0.44
12/06/2013 10:37:52.739018 0.000000094 0.000000996 0.000000770 16.8 12.58 1.15 7996.33 0.49
12/06/2013 10:37:53.739981 0.000000094 0.000000998 0.000000770 16.9 1.14 1.15 9499.65 0.29
12/06/2013 10:37:54.739743 0.000000094 0.000000998 0.000000770 16.9 1.14 1.15 10997.48 0.59
12/06/2013 10:37:55.739858 0.000000094 0.000000996 0.000000770 16.9 14.87 1.15 11999.69 0.42
12/06/2013 10:37:56.740534 0.000000094 0.000001010 0.000000770 16.9 14.87 1.15 13499.35 0.47
12/06/2013 10:37:57.740392 0.000000094 0.000000998 0.000000765 16.9 3.43 1.15 15000.83 0.49
12/06/2013 10:37:58.740247 0.000000094 0.000000998 0.000000768 16.8 3.43 1.15 16499.88 0.57
12/06/2013 10:37:59.740213 0.000000094 0.000000998 0.000000768 16.8 12.58 1.15 18001.98 0.63
12/06/2013 10:38:00.740039 0.000000093 0.000000998 0.000000768 16.8 12.58 1.15 19068.89 0.65
12/06/2013 10:38:01.739890 0.000000093 0.000000998 0.000000774 16.8 10.29 0.97 20003.35 0.65
12/06/2013 10:38:02.739822 0.000000093 0.000000998 0.000000770 16.8 10.29 0.97 20003.35 0.65
12/06/2013 10:38:03.740608 0.000000093 0.000000998 0.000000770 16.8 1007.09 4.08 20005.18 0.65
12/06/2013 10:38:04.740535 0.000000094 0.000000998 0.000000768 16.8 2009.61 2.38 20005.18 0.65
12/06/2013 10:38:05.740421 0.000000094 0.000000998 0.000000768 16.8 3292.51 4.33 20005.18 0.65
12/06/2013 10:38:06.741299 0.000000093 0.000000998 0.000000770 16.8 4510.18 7.87 20005.18 0.65
12/06/2013 10:38:07.741154 0.000000094 0.000000998 0.000000770 16.8 5794.23 7.01 20005.79 0.65
12/06/2013 10:38:08.740978 0.000000094 0.000000994 0.000000770 16.8 7014.19 10.80 20005.79 0.65
12/06/2013 10:38:09.740907 0.000000093 0.000000998 0.000000768 16.8 8294.80 13.91 20005.79 0.65
12/06/2013 10:38:10.741753 0.000000093 0.000000998 0.000000768 16.8 10017.16 14.58 20005.79 0.65
12/06/2013 10:38:11.742650 0.000000094 0.000000996 0.000000770 16.8 11519.79 19.28 20005.79 0.65
12/06/2013 10:38:12.742528 0.000000093 0.000000996 0.000000770 16.8 12021.05 12.81 20005.79 0.65
12/06/2013 10:38:13.742395 0.000000093 0.000000998 0.000000770 16.8 12522.31 12.08 20005.79 0.65
12/06/2013 10:38:14.742226 0.000000093 0.000000998 0.000000770 16.8 13021.28 12.26 20005.79 0.65
12/06/2013 10:38:15.742200 0.000000094 0.000000998 0.000000768 16.8 13801.78 12.57 20005.79 0.65
12/06/2013 10:38:16.742099 0.000000094 0.000000998 0.000000770 16.8 14522.77 17.39 20005.79 0.65
12/06/2013 10:38:17.742891 0.000000093 0.000000996 0.000000768 16.8 15024.03 15.80 20005.79 0.65
12/06/2013 10:38:18.742772 0.000000093 0.000000998 0.000000768 16.8 15524.14 14.89 20005.79 0.65
12/06/2013 10:38:19.742602 0.000000093 0.000000998 0.000000768 16.8 16524.37 18.73 20005.79 0.65
12/06/2013 10:38:20.742477 0.000000093 0.000000998 0.000000768 16.8 17026.77 17.27 20005.79 0.65
12/06/2013 10:38:21.743352 0.000000093 0.000000996 0.000000770 16.8 18030.44 20.81 20005.79 0.65
12/06/2013 10:38:22.743286 0.000000093 0.000000996 0.000000770 16.8 18527.12 18.67 20005.79 0.65
12/06/2013 10:38:23.743116 0.000000093 0.000000998 0.000000770 16.8 19528.49 22.64 20005.79 0.65
12/06/2013 10:38:24.743988 0.000000093 0.000000996 0.000000765 16.8 20029.75 21.17 20005.79 0.65
12/06/2013 10:38:25.743892 0.000000093 0.000000996 0.000000768 16.8 20529.86 20.38 20005.79 0.65
12/06/2013 10:38:26.744754 0.000000093 0.000000998 0.000000770 16.8 21530.09 25.63 20005.79 0.65
12/06/2013 10:38:27.744640 0.000000094 0.000000998 0.000000768 16.8 22031.35 23.49 20005.79 0.65
12/06/2013 10:38:28.744504 0.000000093 0.000000998 0.000000768 16.8 22532.61 22.94 20005.79 0.65
12/06/2013 10:38:29.745410 0.000000094 0.000000998 0.000000768 16.8 23032.73 22.70 20005.79 0.65
12/06/2013 10:38:30.745262 0.000000094 0.000000998 0.000000770 16.8 23533.98 27.52 20005.79 0.65
12/06/2013 10:38:31.745131 0.000000094 0.000000996 0.000000772 16.8 24533.07 26.36 20005.79 0.65
12/06/2013 10:38:32.745008 0.000000094 0.000000998 0.000000768 16.7 25034.33 25.63 20005.79 0.65
12/06/2013 10:38:33.745050 0.000000093 0.000000998 0.000000770 16.7 26034.56 29.90 20005.79 0.65
12/06/2013 10:38:34.744755 0.000000094 0.000001010 0.000000768 16.7 26535.82 28.25 20005.79 0.65
12/06/2013 10:38:35.744622 0.000000093 0.000000998 0.000000770 16.7 27035.93 27.34 20005.79 0.65
12/06/2013 10:38:36.745477 0.000000093 0.000000998 0.000000768 16.7 27989.24 31.73 20005.79 0.65
12/06/2013 10:38:37.746375 0.000000093 0.000000998 0.000000768 16.7 28536.27 30.94 20005.79 0.65
12/06/2013 10:38:38.746259 0.000000094 0.000000998 0.000000770 16.7 29037.53 30.88 20005.79 0.65
12/06/2013 10:38:39.746094 0.000000094 0.000000996 0.000000770 16.7 29538.79 30.51 20005.79 0.65
12/06/2013 10:38:40.746006 0.000000093 0.000000998 0.000000768 16.7 30038.91 30.45 20005.79 0.65
12/06/2013 10:38:41.745828 0.000000093 0.000000998 0.000000770 16.7 30539.02 30.70 20005.79 0.65
12/06/2013 10:38:42.745708 0.000000093 0.000000994 0.000000768 16.7 31040.28 31.55 20005.79 0.65
12/06/2013 10:38:43.746645 0.000000094 0.000000994 0.000000770 16.7 31993.59 36.07 20005.79 0.65
12/06/2013 10:38:44.746538 0.000000093 0.000000998 0.000000765 16.7 32540.62 35.82 20005.79 0.65
12/06/2013 10:38:45.746327 0.000000093 0.000000998 0.000000765 16.7 33040.74 35.15 20005.79 0.65
12/06/2013 10:38:46.747194 0.000000093 0.000001010 0.000000770 16.7 33540.85 34.79 20005.79 0.65
12/06/2013 10:38:47.747076 0.000000093 0.000000996 0.000000770 16.7 34545.66 38.81 20005.79 0.65
12/06/2013 10:38:48.746949 0.000000093 0.000000998 0.000000768 16.7 35042.34 37.17 20005.79 0.65
12/06/2013 10:38:49.747854 0.000000094 0.000000998 0.000000770 16.8 36041.42 46.81 20005.79 0.65
12/06/2013 10:38:50.747794 0.000000093 0.000000996 0.000000768 16.8 36543.83 40.03 20005.79 0.65
12/06/2013 10:38:51.748607 0.000000094 0.000000998 0.000000770 16.8 37043.94 39.61 20005.79 0.65
12/06/2013 10:38:52.748492 0.000000093 0.000000998 0.000000770 16.8 37545.20 39.73 20005.79 0.65
12/06/2013 10:38:53.748395 0.000000093 0.000000998 0.000000768 16.8 38046.46 44.06 20005.79 0.65
12/06/2013 10:38:54.748195 0.000000093 0.000000996 0.000000768 16.8 39044.40 43.03 20005.79 0.65
12/06/2013 10:38:55.748153 0.000000094 0.000000998 0.000000770 16.8 39545.66 43.45 20005.79 0.65
12/06/2013 10:38:56.749112 0.000000094 0.000000998 0.000000770 16.8 39545.66 40.89 20005.79 0.65
12/06/2013 10:38:57.748818 0.000000094 0.000000998 0.000000770 16.8 40046.92 41.50 20005.79 0.65
12/06/2013 10:38:58.748731 0.000000093 0.000000998 0.000000768 16.8 40046.92 41.07 20005.79 0.65
12/06/2013 10:38:59.748612 0.000000094 0.000000996 0.000000768 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:00.748440 0.000000094 0.000000996 0.000000768 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:01.749351 0.000000094 0.000001010 0.000000770 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:02.749233 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.83 20006.40 0.65
12/06/2013 10:39:03.749082 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.83 20006.40 0.65
12/06/2013 10:39:04.749026 0.000000093 0.000000998 0.000000765 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:05.749899 0.000000094 0.000000998 0.000000768 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:06.750723 0.000000093 0.000000998 0.000000770 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:07.750716 0.000000093 0.000000998 0.000000770 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:08.750436 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.83 20005.79 0.65
12/06/2013 10:39:09.750420 0.000000093 0.000000998 0.000000770 16.8 40046.92 41.32 20005.79 0.65
12/06/2013 10:39:10.750265 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:11.751047 0.000000093 0.000000996 0.000000768 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:12.750973 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:13.751821 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:14.751678 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:15.751659 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:16.752539 0.000000094 0.000000998 0.000000772 16.8 40048.06 40.71 20503.85 0.71
12/06/2013 10:39:17.753413 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 21004.34 0.71
12/06/2013 10:39:18.753168 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 21504.23 0.71
12/06/2013 10:39:19.753173 0.000000093 0.000000998 0.000000772 16.7 40048.06 40.58 22070.64 0.76
12/06/2013 10:39:20.752945 0.000000093 0.000000994 0.000000768 16.7 40048.06 40.58 23000.83 0.99
12/06/2013 10:39:21.753831 0.000000094 0.000000996 0.000000772 16.7 40048.06 40.58 23503.16 0.80
12/06/2013 10:39:22.753731 0.000000093 0.000000996 0.000000772 16.7 40048.06 40.58 24003.66 0.80
12/06/2013 10:39:23.753588 0.000000093 0.000000996 0.000000770 16.7 40048.06 40.58 24501.71 0.85
12/06/2013 10:39:24.754474 0.000000094 0.000000998 0.000000765 16.7 40048.06 40.58 25003.43 0.87
12/06/2013 10:39:25.754339 0.000000094 0.000000998 0.000000770 16.7 40048.06 40.71 25503.31 0.87
12/06/2013 10:39:26.754168 0.000000094 0.000000998 0.000000772 16.7 40048.06 40.71 25953.15 0.87
12/06/2013 10:39:27.754087 0.000000093 0.000000998 0.000000772 16.8 40048.06 40.71 26003.20 0.87
12/06/2013 10:39:28.754026 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 26503.70 0.99
12/06/2013 10:39:29.753829 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 26572.67 0.99
12/06/2013 10:39:30.753655 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 27503.47 0.95