Message ID: 421     Entry time: Thursday, August 15, 2013, 16:35
Author: Anders Mjos 
Category: ITE 
Type: Repair 
Module: TM4 
Target/Number: SiC#29 
Subject: Leak found on Coil 1 

By pressurizing the water lines with 70 psi helium David and I detected a major leak on the Coil 1 circuit. The response maxed out on the leak rate on the leak detector and we could also see a pressure increase on the leak detector. ISAC controls reported a change in vacuum at the same time (link).

 

The Heat Shield and Coil 2 circuit was checked before we moved to Coil 1. No response was detected on these two circuits. We would like to check the remaining circuits as well, but this cannot be done today due to the large amount of helium still present. The remaining circuits will have to be checked tomorrow morning.

 

Attachment 1: ITE_Leak_Test_2013-08-14.pdf  80 kB  | Show | Hide all