Message ID: 348     Entry time: Wednesday, May 15, 2013, 16:51
Author: Bevan Moss 
Category: Conditioning Station 
Type: Standard Operation 
Module: TM3 
Target/Number: No Source Tray 
Subject: Repeat of HV Test 

 The HV test completed on Monday was repeated today over a much longer time period. By taking longer there was slightly less sparking but once again a maximum voltage of ~44kV was achieved. The  resistivity of the water was much higher during this run so the leakage current was less but still in the >100 uA range. The spark that occurred during the ramp up to the maximum voltage caused the "water resistivity" limit to trip. This was a feed back through the control not the actual water changing as it recovered instantaneously. 

Attachment 1: 150502013_TM3_No_Source_Tray.xlsx  1.632 MB  Uploaded Wednesday, May 15, 2013, 17:54