Message ID: 2580     Entry time: Thursday, May 02, 2024, 15:23
Author: David Wang 
Category: South Hot-Cell 
Type: Development 
Module: TM3 
Target/Number:  
Subject: Leak check on TM3 new containment box window line VCR fittings.  

After new containment box installation on TM3, Chad and I did a helium spray leak check on two VCR fittings on front side water lines. they are leak free.

Helium pressurize window water line to 70 psi for 3minutes. No response on leak detector which is 0.0xe-9 atm.cc/sec.