Message ID: 2520     Entry time: Thursday, January 11, 2024, 13:54
Author: David Wang 
Category: Conditioning Station 
Type: Development 
Module: TM3 
Target/Number:  
Subject: TM3 servic cap air to vacumm seals leak check. 

After Tm3 service tray and service cap installation, TM3 service cap air to vacuum seals leak check has been done at TCS. TCS:IG1 and IG1S are at low E-7 torr range.  leak detector base  0.0XE-9 atm.cc/sec,  base pressure0.0X E-4 torr.  I sprayed helium to all new air to vacuum seals on service cap. No leak on seals. The leak check has been repeated  2 times with same result.   