Message ID: 242     Entry time: Wednesday, February 06, 2013, 14:34
Author: Bevan Moss 
Category: Conditioning Station 
Type: Development 
Module: TM4 
Target/Number: CS Test Target #1 
Subject: 40kV All Common w/o Water Test 

 Today the module was biased to 40kV in the conditioning station. The current draw was 0uA. Sparking started at 27 kV. I have attached the recording from the strip tool of the vacuum while we were testing. After consulting Pierre if we should take the module to a higher voltage it was decided not to. The purpose of this test we to check the interaction between the module, cover, new cooling line, EE and EINZEL cables. This test was considered a success and although there are no anticipated problems care should be taken when taking the module past 40kV for the first time.