Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC  Not logged in ELOG logo
Message ID: 2369     Entry time: Wednesday, September 28, 2022, 13:15
Author: Matthew Gareau 
Category: North Hot-Cell 
Type: Maintenance 
Module:  
Target/Number:  
Subject: topped up the NHC window oil through the oil expansion tank 

Myself and Maico were able to top up the NHC window oil.

The oil was filled through the expansion tank

- the Nitrogen line closed and the Nitrogen pressure valve was disconnected to ensure that no oil would go in to them.

- the pressure relief valve was disconnected to act as an air bleed line while we filled the oil

- with a funnel we slowly added the oil until the oil level could be read, showing about half full

- about 1.5 liters of oil were added

- once filled the Nitrogen line was reattached and the pressure was returned to recommended level 

- the remaining oil was returned the RH shop

Now the oil level will be monitored, which will help identify if there is any leak in the system.

Attachment 1: IMG_20220928_121000.jpg  3.176 MB  | Show | Hide all
Attachment 2: IMG_20220928_124253.jpg  3.207 MB  | Hide | Hide all
IMG_20220928_124253.jpg
Attachment 3: IMG_20220928_124258.jpg  1.890 MB  | Show | Hide all
ELOG V2.9.2-2455