Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC  Not logged in ELOG logo
Message ID: 1712     Entry time: Wednesday, May 09, 2018, 08:48
Author: David Wang 
Category: ITW 
Type: Repair 
Module:  
Target/Number:  
Subject: 2A2 winodw leak check on ITW vacuum vessel. 

Leak detector is connected to ITW TP4. LD Base leak rate 0.0XE-09 atm.cc/sec. Helium line is connected to existing window leak check 1/8 " copper tube. Applied 18psi helium for 1 minute. No response on LD. The window seal is good on ITW vacuum vessel. 

ELOG V2.9.2-2455