Today a HV test at atmosphere was performed for Each Line(s), while floating the other lines. THIS TEST AT ATMOSPHERE IS THE FIRST ONE AFTER THE FIBER GLASS WAS REMOVED.
Please note:
a). the Einzel Lens line (Y) was grounded.
b). after the fiber glass was removed the "R-U COIL" line and the "(F) Plasma" (which is the Anode line) were shorted to the BIAS voltage.
Tests performed:
1. Applied up to 32kV to "(A) OVEN -"
Sparks occurred at 32kV between the Extraction Electrode holder and the Ground Electrode.
2. Applied up to 33kV to "(B) OVEN +"
Sparks occurred at 33kV but we could not identify the spark location.
3. Applied up to 31kV to "(C) TUBE -"
Sparks occurred at 31kV between the Extraction Electrode holder and the Ground Electrode, at the top and the bottom.
4. Applied up to 30kV to "(D) TUBE +"
Sparks occurred at 30kV between the Extraction Electrode holder and the Ground Electrode, 2 spots at the bottom.
5. Applied up to 30kV to "(P-Q) COIL +"
Sparks occurred at 30kV between the Extraction Electrode holder and the Ground Electrode, all around.
6. Applied up to 30kV to "(R-U) COIL -", to "(F) Plasma" the Anode, to "(E) Cavity" or MSP, "(G) Gas", and "(J) Heat Shield" - all these are "connected" together at the BIAS voltage.
Sparks occurred around 29-30kV at the Extraction Electrode holder and the Ground Electrode, at the bottom.
8. Applied up to 23kV to "(I) EE" - this is for the Extraction Electrode.
Sparks occurred at 23kV again at the same spot near the bottom, between the Extraction Electrode holder and the Ground Electrode.
|