Message ID: 1432     Entry time: Wednesday, February 01, 2017, 12:23
Author: David Wang 
Category: ITE 
Type: Standard Operation 
Module: TM2 
Target/Number:  
Subject: TM2 has been connected in ITE. 

TM2 has been connected in ITE. The SST window water lines in ITE can not be connected properly to TM2. The water line connector on TM2 window is 90 degree CCW installed in direction comparing to TM4.This caused the problem. Solution: Two piece of flexible lines and extra male QC adapters have been used to fix this issue.  HV cover and HV fence in ITE work  with TM2. Other connections are good as well.