Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Evaporator_2  Not logged in ELOG logo
Message ID: 287     Entry time: Thursday, March 24, 2016, 12:05
Author: John Wong 
Target Material: SiC 
Target Oven W.O#:  
Source: TARGET BOAT 
Status: Done 
Subject: Test#3: E-spin SiC with PAN 

Thursday, March 24, 2016, 11:37  

SiC powder (<1um) [10%] was mixed with PAN solution [10%].  Solution was sonicated before e-spinning.  

Fiber mat was binded to graphite foil and heat treated to 250C in air for 1 hour.  

Samples were cut in pieces and rolled up and inserted in Ta target container (TaC coated).  Heat shield was added.

Heat Treatment #1:  

Temperature:  1600C (~350A)

Ramping up TGHT to 300A at 1A/min.  


 Tuesday, March 29, 2016, 11:11

TGHT was maintained at 410A for 1600C for 2 days (see attachment 1).  However at around 22:30 on March 28, there was a sudden increase in voltage.

Temperature raised up to 1750C.  Not sure there was a fault with the control system and it was caused by the Ta-heat shield. (see attachment #2).

 Target was cooled down around 2am on March 29. 

Samples will be removed from the chamber.


 

Attachment 1: Screen_Shot_2016-03-26_at_3.39.35_PM.png  1.211 MB  Uploaded Tuesday, March 29, 2016, 12:12  | Show | Hide all
Attachment 2: TGHT_faulty.PNG  16 kB  Uploaded Tuesday, March 29, 2016, 12:15  | Hide | Hide all
TGHT_faulty.PNG
ELOG V2.9.2-2455