  Evaporator_2  Not logged in ELOG logo
Message ID: 20     Entry time: Tuesday, May 15, 2012, 17:06
Author: John Wong 
Target Material:  
Target Oven W.O#:  
Source:  
Status: Problems resolved 
Subject: TBHT power supply faulty 

 Tuesday, May 15, 2012, 17:03

The tube heater power supply current readback and setpoint are not matching.  Ray Dube checked the power supply and confirmed that it is faulty.  He will replace a functional p/s tomorrow. 


 Friday, May 18, 2012, 13:24

The original p/s was replaced with a new one and there was still a discrepancy between the readback and the setpoint.  After checking with the drawings for the ITE/ITW tube heaters, it was confirmed that the connections for the Evap#2 tube heater control were wrong.  (see attachment 1).  The connections were corrected and the power supply is now operational. 


 

Attachment 1: tubeheater_control.png  834 kB  Uploaded Friday, May 18, 2012, 14:29  | Show | Hide all