Friday, June 13, 2014, 15:02
Target container was previously coated and sintered with TaC. It's wrapped with 3-layers of Ta heat shield, including the center block.
The target consists of 280 discs of UCx with 9cm in length.
A regular TaC coated/sintered endcap is inserted on right, wrt to the ionizer tube pointing down, so the left end is open with no cap to allow out-gassing.
The chamber is being dumping down.
TGHT will be ramped up to 520A - keep heating at that current for ~4 to 5 hours.
Parameter setup: 5.0e-5 good vac, 6.0e-5 bad vac. The auto-ramping rate is set to 0.1A/min.
Friday, June 13, 2014, 16:26
Pressure at 1.9e-5 Torr.
Initiate auto-ramping- stepping up from 0A at 0.1A/min - leaving for the weekend.
Wednesday, June 25, 2014, 11:45
TGHT has reached 520A and target has been conditioned at that current for 13hrs.
Auto-cooling initiated at 2A/min.
Once the target is cooled, the chamber will be vented in Ar.
Friday, July 04, 2014, 10:06
Process is complete. Target is stored in the Ar-glove box and ready for assembly.
|