Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Evaporator_1  Not logged in ELOG logo
Message ID: 146     Entry time: Thursday, April 20, 2017, 11:41
Author: Laura Lambert 
Target Material: UC2 
Target Oven W.O#: 42218-1 
Source: LP-IG-LIS 
Status: Done 
Subject: Conditioning UCx#20 

 Thursday, April 20, 2017, 11:31

Target container was previously coated with TaC spray and sintered.

Container is wrapped with 3 layers of Ta-heat shield, including the center block

The target consists of 220 discs of UCx with 7.55cm in length [6 packs of 5 + 190 single discs].

Discs in target are arranged as follows: 1pack / 45singles / 1pack / 50singles / 2packs / 50singles / 1pack / 45singles / 1pack

A regular TaC coated endcap is inserted left, wrt the ionizer tube pointing down.

The right side is open with a TaC coated spring to help hold materials in.

The chamber is currently pumping down.

TGHT will be ramped up to 490A at 0.1A/min [same current used to reach 1900C in evap2 when sintering TaC coating - elog350 evap2]

Parameter setup: 5.0e-5 good vac, 6.0e-5 bad vac.


 Friday, April 21, 2017, 10:14

IG1 = 2.47e-6 Torr

Tapped up manually to 35A.

Waited ~1.5hrs, vacuum stable IG1 = 2.37e-6 Torr

Initiating auto-ramp.


 Thursday, April 27, 2017, 09:19

IG1 = 4.611e-6 Torr

TGHT = 490A, 4.93V

Target has remained at max setpoints for ~9hrs.

Initiating auto-cool at 1.0A/min


 Tuesday, May 16, 2017, 13:21

Update: 

Target looks good, see photos attached post sintering.

Gold tint on container commonly seen with the Uranium targets.

Orange colouring on inside of container end - also common.

Both endcaps in and handed off to assembly yesterday.


 

Attachment 1: 20170420_103410.jpg  1.137 MB  | Show | Show all
Attachment 2: 20170420_103516.jpg  1.481 MB  | Show | Show all
Attachment 3: 20170420_103154_Richtone(HDR).jpg  1.889 MB  | Show | Show all
Attachment 4: 20170508_150816_Richtone(HDR).jpg  1.546 MB  Uploaded Tuesday, May 16, 2017, 14:23  | Show | Show all
Attachment 5: 20170512_101150_Richtone(HDR).jpg  1.489 MB  Uploaded Tuesday, May 16, 2017, 14:23  | Show | Show all
Attachment 6: 20170512_101348_Richtone(HDR).jpg  1.500 MB  Uploaded Tuesday, May 16, 2017, 14:23  | Show | Show all
ELOG V2.9.2-2455