OLIS Logbook General
  General, Page 25 of 26  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  392   Fri Mar 25 17:42:38 2011 Keerthi Jayamanna  OLIS HV start up

High Voltage cage is on at ~17:43pm

Interlock cleard and HV (10kV) is on at 17:53pm

  415   Wed Jul 20 01:37:46 2011 Keerthi Jayamanna  Al beam development:

 

Supernanogan is tested for Al+6 beam production.
A 8nA of mass 4.5 is seen. Background pressure is not good enough to
produce more Al+6. Attempt to increase Al beam current with higher sputtering
voltages resulted in fast destruction of the Al sample.
Decided to move to the microwave ion source.
Clean all the Mg and installed Al liners and the Al PE electrode.

Source is now pumping down. Damien, Please start the source as soon as you arive. Try to keep PE current below 500mA. Optimum is ~200mA

  417   Wed Aug 3 00:15:39 2011 Keerthi JayamannaOtherGeneral22Ne+2

Source started at 800W of microwave power and 2cc/min neon gas easily. New magnetron produces more power than the old one.

We should try to keep MW power under 500W during running and use only 800W for ignition.  EE left off for the night.

  468   Wed May 23 17:11:57 2012 Keerthi Jayamanna  Changed beam to 12C++ beam

Removed titanium from the back plate and clean the source.

But left the PE titanium insert since Bob and Marco badly want the 12C++ beam from the source fast.

Background 12C++ was only 400pA so installed a small canister with CO in it with bellow atm. pressure.

Now source is running with ~2nA current.

  469   Sun Dec 2 22:32:19 2012 Keerthi JayamannaRoutineGeneralStill cannot write anything other than Subject
  470   Sun Dec 2 22:34:09 2012 Keerthi Jayamanna  Just finished 136Xe+25. ISAC reached 1GeV yesterday using OLIS Xe beam
  471   Sun Dec 2 23:36:13 2012 Keerthi JayamannaRoutineGeneral16O+2 is ready for tuning.
MCIS finished xenon development. MCIS is off and isolation valve is shut.
A new oxygen gas line is installed to microwave ion source via a hand valve neat PE.
Hand valve is tuned as a small leak so that the plasma can run with all other gas valves are shut.
However if one need to start the source Ne gas valves are needed to be opened.
A 3.13nA of 16O+2 is sitting at FC6 and ready to use as pilot beam.
  472   Fri Dec 14 17:30:47 2012 Keerthi Jayamanna  18OH2 and 40Ar+2 delivary

During the last week we delivered 18O+2 from microwave ion source to Ritu.

Over the week end changed to 40Ar+2.

Ritu informed us 18OH2 is coming with 40Ar+2.

Moving the slits ops managed to reduce 18H2 contaminant from the beam.

Good lesson for the future. Bad combination one after the other.

  473   Mon Dec 17 14:08:52 2012 Keerthi Jayamanna  IRIS beam delivary is completed

IRIS beam delivery is completed with extra 3 days of beam providing to them.

18O2 and 40Ar+2 were delivered with few attenuators i place.

  474   Mon Dec 17 14:10:22 2012 Keerthi Jayamanna  Gas flow controller fault

Controls group may need some assistance from OLIS group in order to trouble shoot the problem.

  493   Thu Oct 30 22:09:02 2014 Keerthi Jayamanna  Suoernanogan steering

IOS mass scan does not work.

A/Q            FC6                X          Y          CCB1

1                2.07e-7        167      97       500

2                2.18e-7         308     247

2.14          1e-11             388     277 

2.246       2.07e-8         358     227

2.30          2e-9             378      227 

2.366       1.28e-8         388     277

2.392        2.1e-11        388     277

2.4641      1.98e-11     388     277

2.5345      6.49e-11     398     277

2.5614     1.81e-10      398    227

2.6263     2.95e-7         398   227

2.7522     2.07e-8         398   227

2.7896     2.06e-10       398   227

2.8652     3.22e-11       398   227

2.9515     8.52e-8         398   237

 

 

 

 

 

 

 

 

 

 

 

  494   Thu Oct 30 22:44:45 2014 Keerthi JayamannaRoutineGeneralMass scan 3 to 4

IOS mass scan if faulty. These is a manual scan from mass 3 to 4

Const22 = 0.04187  adjusted to mass = 4

A/q               FC6           X          Y 

 

3.0075        8.14e-8      398  237

3.1222        4.93e-10   398  237

3.2064        6.24e-7     398   237

3.2613        1.05e-9     408   237

3.3395        2.29e-10   408  237

3.4072        5.6e-10     408  237

3.5089        3.36e-8     408  237

3.5600        1.4e-10     408  237

3.6069        1.86-9       408  237

3.6436        8.46e-10  408  237

3.7514        4.02e-10  418   237

3.8387        4.37e-9    418   237

3.8883        3.43e-10  418   237

3.9416        1e-10       418    237

4                   1.1e-6     418    237       500  

  495   Thu Oct 30 23:37:58 2014 Keerthi JayamannaRoutineGeneralMass scan 4 to 6

A/Q            FC6

4                1.1e-6

4.25          8e-10

4.34          1.2e-9

4.42          5.5e-9

4.51          2.59e-9

4.58          1.7e-10

4.66          1.7e-10

4.74          1.7e-10

4.85          4.6e-10

5.0            1.36e-10

5.07          1.5e-10

5.13         2.69e-10

5.25         4.89e-10

5.33         9e-7

5.39        3.0e-10

5.56        3.3e-10

5.67        1.12e-9

5.7          5e-9

5.83       4e-10

5.92       2e-10

6.000     1.85e-7

 

  496   Thu Oct 30 23:44:57 2014 Keerthi JayamannaRoutineGeneralOLIS Supernanogan mass scan 6 to 7.5

A/q           FC6  (only current over 100pA)

6.0           1.85e-7

6.15         4.96e-10

6.23         2.4e-10

6.31         6.14e-10

6.39         2e-10

6.51         1.82e-9

6.90         2.7e-10

7.00         5e-8

7.1           4.6e-9

7.27        1e-10

7.5        3.32e-10

 

  498   Sat Nov 1 13:31:17 2014 Keerthi Jayamanna  MCIS readyness for 76Se

Supernanogan is running with oven no heat, without 76selenium. MCIS:IG1=3e-7Torr.

Tune saved as 141101_1318_76Se+11tune.snap

MCIS steering  X=330, Y=226, CCB1=500, MCISEL=4171/5157, PE=11.5/23, FC6=76nA(16O+7)

  506   Tue Nov 25 12:23:28 2014 Keerthi JayamannaRoutineGeneral76Se+13 tune saved

141125_1159_OLIS_76Se+13.snap is saved in reference directory.

El=3480/4303V

MCIS:YCB1=110

MCIS:XCB1=190

MCIS:CCB1=300

Frequency = 14.44765GHz

Amplitude = -10.880

TET FP= 9.8W

TWT RP = 1.8W

 

  511   Mon Sep 14 11:21:16 2015 Keerthi JayamannaRoutineGeneral20Ne+2 beam saved

20Ne+2 beam optics saved as 150914_1117_ios0iosfc6.snap

  512   Wed Feb 22 21:46:27 2017 Keerthi Jayamanna  Elog is working now

Elog is now functioning. We can start entering notes

  513   Wed Feb 22 21:51:40 2017 Keerthi Jayamanna  SIS first beam is N1+1

Chris confirmed the first SIS beam would be Na+1. Next Rb+1 and then Li+1

  484   Fri Oct 10 14:02:56 2014 Raymond DubeRoutineGeneralTest with new Alga, ALPA 50W RF AMP

We did a test yesterday with the new Alga Microwave RF AMP, model # ALPA 50 Watt, serial # 130684 to be used for the MCIS.  We used the RF Generator from MISTIC and set the frequency to 13.80 Ghz, note that the RF Generator was/is  limited to 16DB maxiumum RF output.

The RF cable that we used was from the 200W RF AMP, it has a 20DB SMA attenuator at the Generator end, so we could only get about 25 Watts maximum output.

Attachment 1: ALPA-RF_AMP-50W.pdf
ELOG V2.9.2-2455