ID Date Author Type Categoryup Subject
  51   Wed Jul 8 11:52:54 2009 Damien GallopRoutineGeneralSupernanogan development run plan for the remainder of this set

* 18On+ emittance scans at present ion source settings: complete today.

* 18On+ frequency measurements: Thursday.

* 17O source gas: open end of Thursday.

* 17On+ frequency survey: Friday.

* 17On+ emittance scans at idealized source settings: Monday.

* Development completion: Tuesday.

* 18O4+ setup for beam schedule: Wednesday.

  52   Wed Jul 8 16:12:51 2009 Damien GallopRoutineGeneralMass scan in progress

A mass scan from A/q of zero to 25 is in progress. Data file (large): masscorr_mcis_090708.txt. The source is optimized for 18O4+. Fishing for 18On+ states and their emittance scans to follow. Attached is one screenshot of the Gabe's Calibrate interface. I ran it several times with different peak selections from the same data. I entered into the EPICS mass correlation program a best average of the calibrations for the two constants.

Attachment 1: grab.rh2790.png
grab.rh2790.png
Attachment 2: Picture_16.png
Picture_16.png
Attachment 3: Picture_17.png
Picture_17.png
  53   Wed Jul 8 17:47:26 2009 Damien GallopRoutineGeneral18O4+ emittance scans

First up is a check scan of 18O4+. There is a definite vertical offset compared to earlier today. Otherwise, it correlates well enough. The same beam taken with a greater vertical offset follows, together with the modified emittance panel.

Attachment 1: grab.Jj2874.png
grab.Jj2874.png
Attachment 2: grab.fz2878.png
grab.fz2878.png
Attachment 3: grab.ND2931.png
grab.ND2931.png
Attachment 4: grab.pH2935.png
grab.pH2935.png
Attachment 5: grab.oY2939.png
grab.oY2939.png
  54   Wed Jul 8 17:58:02 2009 Damien GallopRoutineGeneral18O1+ emittance scan

Here we go. A/q = 18 FC6 = 141.0nA MB = 2078G @ 130.2A

Attachment 1: grab.QR2996.png
grab.QR2996.png
Attachment 2: grab.dm3000.png
grab.dm3000.png
  55   Wed Jul 8 18:12:25 2009 Damien GallopRoutineGeneral18O2+ emittance scan

A/q = 9 FC6 = 50.0nA MB = 1470.8G @ 90.8A

Attachment 1: grab.LL3034.png
grab.LL3034.png
Attachment 2: grab.Yj3038.png
grab.Yj3038.png
  56   Wed Jul 8 18:20:43 2009 Damien GallopRoutineGeneral18O3+ emittance scan

A/q = 6 FC6 = 49.5nA MB = 1201.4G @ 73.1A

Attachment 1: grab.Pc3095.png
grab.Pc3095.png
Attachment 2: grab.Oi3099.png
grab.Oi3099.png
  57   Wed Jul 8 18:38:21 2009 Damien GallopRoutineGeneral18O4+ emittance scan

A/q = 4.5 FC6 = 29.5nA MB = 1041.8G @ 62.4A

Attachment 1: grab.Cb3157.png
grab.Cb3157.png
Attachment 2: grab.wO3161.png
grab.wO3161.png
  58   Wed Jul 8 18:47:43 2009 Damien GallopRoutineGeneral18O5+ emittance scan

A/q = 3.6 FC6 = 12.5nA MB = 931.8G @ 55.3A

Attachment 1: grab.IW3188.png
grab.IW3188.png
Attachment 2: grab.kW3192.png
grab.kW3192.png
  59   Wed Jul 8 18:57:09 2009 Damien GallopRoutineGeneral18O6+ emittance scan

A/q = 3 FC6 = 5.5nA MB = 851.5G @ 50.1A

Attachment 1: grab.cb3228.png
grab.cb3228.png
Attachment 2: grab.un3232.png
grab.un3232.png
  60   Wed Jul 8 19:07:15 2009 Damien GallopRoutineGeneral18O7+ emittance scan

A/q = 2.57 FC6 = 0.25nA MB = 790.0G @ 46.2A

Not enough intensity for an emittance scan.

Attachment 1: grab.ca3306.png
grab.ca3306.png
  61   Wed Jul 8 19:13:15 2009 Damien GallopRoutineGeneral18O8+ emittance scan

A/q = 2.25 FC6 at reverse bias MB = 736.9G @ 42.9A

Not enough intensity for an emittance scan. Possibly on the edge of A/q=2.3 (23Na10+?).

Attachment 1: grab.Ah3293.png
grab.Ah3293.png
Attachment 2: grab.ed3302.png
grab.ed3302.png
Attachment 3: grab.Nc3298.png
grab.Nc3298.png
  62   Wed Jul 8 19:51:33 2009 Damien GallopRoutineGeneral18On+ summary for the day

The 18O source gas bottle is online. The ion source was optimized for 18O4+ and kept the same throughout. Emittance scans of each charge state of 18O were attempted. Visible scans were found for all states up to 6+. At least one more set of scans is desireable. Present beam is 18O4+. Next up: RF frequency survey with 18On+.

Attachment 1: grab.TF3397.png
grab.TF3397.png
Attachment 2: grab.fu3402.png
grab.fu3402.png
Attachment 3: grab.Tk3406.png
grab.Tk3406.png
  63   Thu Jul 9 16:07:56 2009 Damien GallopRoutineGeneral18O8+ frequency scans

Keerthi did some 18On+ frequency scans this morning. He observed 18O8+! I am attempting to repeat it myself now. Reverse bias on FC6.

Peaks at:

13.0277GHz => 1epA (unstable)

13.1023GHz => 1epA (unstable)

13.1695GHz => 10epA (unstable)

13.1672GHz => 20epA (unstable) stripchart attached

13.006GHz => .7epA stripchart at 17:24 attached

First, optimize current versus frequency and amplitude. Unfortunately at 5dB the beam blinked out. Go fishing, then find it again. Attached is stripchart at 17:30, where I was optimizing, then switched bias to Normal, then watched beam blink out. Then stripchart at 17:49 shows beam back again, reverse bias on the left and normal bias on the right: ~10epA. RF settings also attached. Minutes later, beam is still on. FTR, the original RF is appended at the end.

 

Attachment 1: grab.hM6579.png
grab.hM6579.png
Attachment 2: grab.OC6749.png
grab.OC6749.png
Attachment 3: grab.pj6791.png
grab.pj6791.png
Attachment 4: grab.Zw6832.png
grab.Zw6832.png
Attachment 5: grab.ko6836.png
grab.ko6836.png
Attachment 6: grab.xu6426.png
grab.xu6426.png
  64   Fri Jul 10 09:55:25 2009 Damien GallopRoutineGeneral18O8+ beam overnight

Here is a striptool of 18O8+ beam (magenta trace) last evening from about 18:00 to 22:30 when there was a large power bump coincident with a plane crash in Richmond last night. Before the trip, beam current was about 20epA on FC6 (normal bias).

Attachment 1: grab.N10241.png
grab.N10241.png
  65   Fri Jul 10 11:46:55 2009 Damien GallopRoutineGeneral18On+ frequency scans & resonances

The Supernanogan was restarted this morning, after last night's power bump. 18O4+ on FC6, cooking for about an hour. Looks very good. Keerthi is doing frequency scans today to find best resonances. Time permitting, he will do this over a range of 18O charge states. At some point over the weekend (my guess is tomorrow evening), he will open the 17O source gas bottle to the system.

Attachment 1: Picture_18.png
Picture_18.png
  66   Fri Jul 10 11:47:07 2009 Damien GallopRoutineGeneral18On+ frequency scans & resonances

The Supernanogan was restarted this morning, after last night's power bump. 18O4+ on FC6, cooking for about an hour. Looks very good. Keerthi is doing frequency scans today to find best resonances. Time permitting, he will do this over a range of 18O charge states. He is also developing a dual-mode (MCIS/MWS) IOS tune. At some point over the weekend (my guess is tomorrow evening), he will open the 17O source gas bottle to the system.

Attachment 1: Picture_18.png
Picture_18.png
  133   Fri Sep 11 23:42:45 2009 KeerthiOtherGeneralMCIS turbo pump

MCIS TP1 is not running well. Needs investigation as soon as S1219 is completed.

Later... TP1 failed completely, and we are running without it. Starting late Saturday night, DRAGON began to observe a certain instability. We correlated it to MCIS:COL0, which skims total current continuously. It is pretty evident that the source is now toggling between two adjacent plasma modes, due to the somewhat higher vacuum at MCIS. Vacuum was something in the 7-8e-8 Torr range before, but has settled out at some 1.25e-7 Torr now. General consensus is to continue running like this, as it is a minor effect, as you can see below (blue trace). DMG

Attachment 1: mcis_multi-state.png
mcis_multi-state.png
  138   Tue Sep 15 19:46:21 2009 KeerthiRoutineGeneralSupernanogan emittance scans

090915_1915ILTEMIT.txt = 17O +3,  4.4uA new tune, scanner offset +10V,  scanner scale 100nA , 100x50 matrix

090915_1929ILTEMIT.txt = same as above with 200x100 matrix

090915_1949ILTEMIT.txt = after tune tweaking (not optimal; for reference only)

Attachment 1: grab.w23925.png
grab.w23925.png
Attachment 2: grab.L23930.png
grab.L23930.png
Attachment 3: grab.J23942.png
grab.J23942.png
Attachment 4: grab.K23948.png
grab.K23948.png
Attachment 5: grab.N23902.png
grab.N23902.png
Attachment 6: grab.f23908.png
grab.f23908.png
Attachment 7: grab.C23914.png
grab.C23914.png
Attachment 8: grab.v23919.png
grab.v23919.png
Attachment 9: grab.y24035.png
grab.y24035.png
  140   Wed Sep 16 12:12:04 2009 KeerthiRoutineGeneral17O+3 emittance sacns

090916_1156ILTEMIT.txt      17O+3  good settings 50x100 ,   333nA Scale, ~4uA

090916_1206ILTEMIT.txt      17O+3  good settings 100x200 ,   333nA Scale, ~4uA

  142   Wed Sep 16 12:26:51 2009 KeerthiRoutineGeneral17O+3 emittance with changed steering

090916_1222ILTEMIT.txt  Changed the MCIS steering to center the beam in the EL. 17O+3 ~5uA, 50x100