Entry  Wed Jul 8 17:47:26 2009, Damien Gallop, Routine, General, 18O4+ emittance scans grab.Jj2874.pnggrab.fz2878.pnggrab.ND2931.pnggrab.pH2935.pnggrab.oY2939.png
First up is a check scan of 18O4+. There is a definite vertical offset compared to earlier today. Otherwise, it correlates well enough. The same beam
taken with a greater vertical offset follows, together with the modified emittance panel.
Entry  Wed Jul 8 18:38:21 2009, Damien Gallop, Routine, General, 18O4+ emittance scan grab.Cb3157.pnggrab.wO3161.png
A/q = 4.5 FC6 = 29.5nA MB = 1041.8G @ 62.4A
Entry  Wed Jul 8 09:22:07 2009, Damien Gallop, Routine, General, 18O4+ beam growth overnight grab.hD1461.png
After accessing the OLIS cage and opening the 18O source gas valve, I restarted the Supernanogan source last evening around 18:00 ( just to the left
of -15h on the attached stripchart), with the line set up for 18O4+. The beam on FC6 grew nicely overnight. MCIS pressure moved from ~6.5e-8 to ~7.1e-8
Torr since the 18O source gas bottle was opened last evening.
Entry  Thu Jul 16 16:33:28 2009, Damien Gallop, , , 18O3+ survey grab.k17100.pnggrab.e17104.pnggrab.E17109.png
A/q = 6 Field = 1147.2 FC6 = 39.3enA
Entry  Wed Jul 8 18:20:43 2009, Damien Gallop, Routine, General, 18O3+ emittance scan grab.Pc3095.pnggrab.Oi3099.png
A/q = 6 FC6 = 49.5nA MB = 1201.4G @ 73.1A
Entry  Thu Jul 16 16:50:41 2009, Damien Gallop, , , 18O2+ survey grab.N17330.pnggrab.Y17335.pnggrab.w17339.png
A/q = 9 Field = 1405.0G FC6 = 55enA
Entry  Wed Jul 8 18:12:25 2009, Damien Gallop, Routine, General, 18O2+ emittance scan grab.LL3034.pnggrab.Yj3038.png
A/q = 9 FC6 = 50.0nA MB = 1470.8G @ 90.8A
Entry  Thu Jul 16 17:09:52 2009, Damien Gallop, , , 18O1+ survey grab.b17471.pnggrab.r17475.pnggrab.R17480.png
A/q = 18 Field = 1987.1 FC6 = 161nA
Entry  Wed Jul 8 17:58:02 2009, Damien Gallop, Routine, General, 18O1+ emittance scan grab.QR2996.pnggrab.dm3000.png
Here we go. A/q = 18 FC6 = 141.0nA MB = 2078G @ 130.2A
Entry  Thu Jun 4 10:46:37 2009, Damien Gallop, Routine, General, 18O source operation overnight 
The MWS delivered 18O overnight, mostly onto FC6. Sophia complained about source stability this morning. It definitely could be improved. Yesterday Keerthi
did some online tweaking of the RF power for stability, with little improvement. The IOS pressure today is about 2e-6 Torr, so I believe the 18O pressure
is too high for good operation, although the brightness is great. The base pressure is about 4e-7 Torr. Tuning is in progress. Given an opportunity, I
Entry  Mon Jul 6 17:34:24 2009, Damien Gallop, Routine, General, 18O background studies 6x
Well... my browser blew away just before I submitted version one of this entry. Let's try again.
Keerthi changed the tune this morning to 10kV on the MCIS bias in order to allow the full throw of the Enzel lens power supply. (Remember, for
the moment the panel displays twice what the physical power supply is actually delivering, because it is a 10kV supply, whereas the control system expects
Entry  Tue Sep 15 20:02:46 2009, Keerthi, , , 17O8+ on FC6 from MCIS grab.I24583.png
0.4nA of 17O+8 is sitting on FC6. Development will continue tomorrow.
Entry  Fri Nov 6 11:17:54 2009, Damien Gallop, , , 17O3+ setup to DRAGON Picture_9.png
As noted, we handed over the beam to our users yesterday by noon. Setup to DRAGON is continuing. The report this morning that beam is great. Good.
Entry  Fri Aug 19 10:01:52 2011, Damien Gallop, , , 17O3+ from MCIS to DRAGON Picture_20.pngPicture_21.png
Setup in progress. No foil at M:5. Source has been running since yesterday. Note that source was opened recently for MRO. Vacuum pumpdown chart for past
several days attached.
Later... ~10:00 ICR reported MCIS:PV1 closed. On closer inspection, it was pretty clear that the classic PLC "all PVs go to zero" state
Entry  Mon Aug 22 09:49:34 2011, Damien Gallop, , , 17O3+ from MCIS to DRAGON Picture_25.png
Delivery continues. The attached vacuum trar suggests that this is about as good as it's going to get. On the other hand, there is a uA of beam
on FC6.
Entry  Wed Aug 24 10:49:41 2011, Damien Gallop, , , 17O3+ from MCIS to DRAGON Picture_20.pngPicture_21.pngPicture_22.png
Delivery continues. There were several "PV1" trips over the past several days. ICR now has a recipe for resetting this condition. We await
a general instruction from ICR via email.
Meanwhile, MCIS:TP3 was off for the last while. ICR notified. Investigating. Turned on again.
Entry  Thu Aug 25 10:54:34 2011, Damien Gallop, , , 17O3+ from MCIS to DRAGON Picture_24.pngPicture_25.pngPicture_26.pngPicture_27.pngPicture_28.png
Delivery continues.
Entry  Mon Aug 29 10:03:05 2011, Damien Gallop, , , 17O3+ from MCIS to DRAGON 
Delivery continues on autopilot. Looking good.
ICR noted that a seemingly unusual number of tweaks to the IOS separator magnet have been required. This is not so surprising, given that the magstab program
is not running presently. It is expected that today will be EOB for DRAGON. A switch to 58Ni is on deck.
Entry  Tue Aug 30 10:01:30 2011, Damien Gallop, , , 17O3+ from MCIS Reserved Picture_32.png
We are on orders to retain OLIS for accelerator development. The switchover to 58Ni, scheduled for this morning, is on hold.
Meanwhile, there were WFB1 and PV1 trips this morning. All reset and beam is on. The WFB paddle for the MCISRFS circuit was still rotating at a
reasonable rate. I adjusted the WFB setting for ON state with minimum flow.
Entry  Mon Nov 16 10:00:41 2009, Damien Gallop, , , 17O3+ beam to DRAGON end of run this morning Picture_1.png
DRAGON declared EOB at around 08:00 today. Source and beam were switched to 21Ne4+ for calibration studies.