Entry  Thu Dec 3 14:12:56 2009, Damien Gallop, , , 20Ne1+ at 28keV for 8pi delivery continues Picture_20.png
Beam was idle overnight. We took the beam for a while this morning, and ran a mass scan. We opted to not go into the cage to install helium gas lines
or fix water lines, because it would mean switching off the source. So we will do the gas tomorrow or so, and fix the water leaks then too.
Note that the separator field is up about 4G from this morning.
Entry  Fri Dec 4 09:58:08 2009, Damien Gallop, , , 20Ne1+ at 28keV for 8pi delivery complete grab.InU319.pnggrab.YPZ323.pnggrab.RXD329.pnggrab.aGj522.pngPicture_1.png
Mike declared successful EOB this morning. "Best beam I ever used" or something like that! EOB emittance and mass scans attached. Note that
slits are at non-standard positions for 5mm gap.
Now we are going to open the source for some maintenance and continue 4He2+ development after that.
Entry  Wed Dec 2 19:03:55 2009, Damien Gallop, , , 20Ne1+ at 28keV for 8pi delivery Picture_19.png
The source ran well overnight and today. There is an issue with an ILE Faraday cup. Tuning is ended for today, and beam is on FC6.
Entry  Tue Dec 1 12:33:58 2009, Damien Gallop, , , 20Ne1+ at 28keV for 8pi 8x
By agreement, we are providing OLIS beam to the Beam Delivery group for the purpose of checking out transmission in the 8pi line. We have set up a 20Ne1+
beam at 28keV. There are some 3nA on FC6.
Here are optics numbers for IOS section:
Entry  Mon Sep 14 11:21:16 2015, Keerthi Jayamanna, Routine, General, 20Ne+2 beam saved 
20Ne+2 beam optics saved as 150914_1117_ios0iosfc6.snap
Entry  Fri May 15 14:12:45 2009, Geoff Wight, , , 1st test 
test
Entry  Thu Oct 6 15:15:09 2011, Damien Gallop, , , 19F/40Ar from MCIS for Beam Development Picture_4.png
Slit calibration exercise was wound up this morning. MCIS source gases were reconfigured with 19F and 40Ar by request. A "cocktail" beam will
be used for HE development over the Thanksgiving weekend. A mass scan is in progress presently.
Entry  Mon Oct 17 11:33:03 2011, Damien Gallop, , , 19F/38Ar from MCIS for Beam Development 
Source set up for a cocktail beam since yesterday, Sunday.
Entry  Tue Oct 18 10:27:44 2011, Damien Gallop, , , 19F/38Ar from MCIS for Beam Development Picture_7.png
Cocktail beams continue today.
Entry  Wed Jul 8 19:51:33 2009, Damien Gallop, Routine, General, 18On+ summary for the day grab.TF3397.pnggrab.fu3402.pnggrab.Tk3406.png
The 18O source gas bottle is online. The ion source was optimized for 18O4+ and kept the same throughout. Emittance scans of each charge state of 18O
were attempted. Visible scans were found for all states up to 6+. At least one more set of scans is desireable. Present beam is 18O4+. Next up: RF frequency
survey with 18On+.
Entry  Fri Jul 10 11:46:55 2009, Damien Gallop, Routine, General, 18On+ frequency scans & resonances Picture_18.png
The Supernanogan was restarted this morning, after last night's power bump. 18O4+ on FC6, cooking for about an hour. Looks very good. Keerthi is doing
frequency scans today to find best resonances. Time permitting, he will do this over a range of 18O charge states. At some point over the weekend (my guess
is tomorrow evening), he will open the 17O source gas bottle to the system.
Entry  Fri Jul 10 11:47:07 2009, Damien Gallop, Routine, General, 18On+ frequency scans & resonances Picture_18.png
The Supernanogan was restarted this morning, after last night's power bump. 18O4+ on FC6, cooking for about an hour. Looks very good. Keerthi is doing
frequency scans today to find best resonances. Time permitting, he will do this over a range of 18O charge states. He is also developing a dual-mode (MCIS/MWS)
IOS tune. At some point over the weekend (my guess is tomorrow evening), he will open the 17O source gas bottle to the system.
Entry  Tue Jul 7 17:47:14 2009, Damien Gallop, Routine, General, 18On+ beam & emittance 
As noted earlier, we have 17O and 18O source gas attached to the Supernanogan source chamber. Both are valved off presently. The source has been running
for a while. I set it to 18O4+ on background, which gives about 125epA on FC6. Now let's stop the source, open the bottle and try the same again.
I opened the 18O valve just around 18:00. Beam is on now, around 150pA. That's a good sign. It's growing too. Even better. The MCIS tank pressure
Entry  Fri Dec 14 17:30:47 2012, Keerthi Jayamanna, , , 18OH2 and 40Ar+2 delivary 
During the last week we delivered 18O+2 from microwave ion source to Ritu.
Over the week end changed to 40Ar+2.
Ritu informed us 18OH2 is coming with 40Ar+2.
Entry  Wed Jul 15 10:41:23 2009, Damien Gallop, , , 18O8+ survey grab.dg7993.png
A/q = 2.25 Field = 704.7G FC6 = <pA
Well, it's there, but pretty faint. The attached stripchart shows no beam then beam. FC6 bias has to be revered to see it.
Entry  Thu Jul 16 14:44:06 2009, Damien Gallop, , , 18O8+ survey 
A/q = 2.25 Field = 702.5G FC6 = not visible
It looks like last night's vacuum trip is biting us here.
Entry  Fri Jul 17 09:47:16 2009, Damien Gallop, , , 18O8+ recheck 
Go fishing in the range 2.24-2.26, but no fish in this pond this time. However, there is a relatively strong beam at 2.3.
A/q = 2.3 Field = 711.2 FC6 = 110enA
No emittance scan possible.
Entry  Thu Jul 9 16:07:56 2009, Damien Gallop, Routine, General, 18O8+ frequency scans 6x
Keerthi did some 18On+ frequency scans this morning. He observed 18O8+! I am attempting to repeat it myself now. Reverse bias on FC6.
Peaks at:
13.0277GHz => 1epA (unstable)
Entry  Wed Jul 8 19:13:15 2009, Damien Gallop, Routine, General, 18O8+ emittance scan grab.Ah3293.pnggrab.ed3302.pnggrab.Nc3298.png
A/q = 2.25 FC6 at reverse bias MB = 736.9G @ 42.9A
Not enough intensity for an emittance scan. Possibly on the edge of A/q=2.3 (23Na10+?).
Entry  Fri Jul 10 09:55:25 2009, Damien Gallop, Routine, General, 18O8+ beam overnight grab.N10241.png
Here is a striptool of 18O8+ beam (magenta trace) last evening from about 18:00 to 22:30 when there was a large power bump coincident with a plane
crash in Richmond last night. Before the trip, beam current was about 20epA on FC6 (normal bias).