The previous 12C2+ development run from the Supernanogan ended Friday morning. Starting Friday afternoon, the Supernanogan was prepared for an upcoming SEBT development using krypton beams. The source was ready overnight. There were some teething pains over the weekend. An A/q=5.71 beam is sitting on FC6 presently. The beam schedule shows an experiment change on Thursday morning. 
Later... Some notes about krypton. 
Naturally occurring isotopes: 
Kr-78   ( 0.35% , 2.0E+21 ) Kr-80   ( 2.25% ) Kr-82   ( 11.6% ) Kr-83   ( 11.5% ) Kr-84   ( 57.0% ) Kr-86   ( 17.3% ) 
Some A/q values: 
    - 40/7 = 5.7143
 
    - 80/14 = 5.7143
 
    - 82/14 = 5.8571
 
    - 82/15 = 5.4667
 
    - 83/15 = 5.5333
 
    - 84/15 = 5.6
 
    - 86/15 = 5.7333
 
 
Keerthi's mass scans from Saturday: 
    - Background scan
 
    - Krypton scan (note multi-peaks in A/q = 5.7 to 5.75 range)
 
    - FC6 at A/q=7.3333
 
 
  
   |