20Ne1+ set up from the MWS and handed over to Sophia & Marco for accelerator recommissioning. Strip to 5+ at M:5. First day of beam. Happy St. Patrick's Day.
Later... a Stripchart of beam today: baby-blue trace is FC6. Also, here are screenshots of the source.
..................................................................................
Lots happened during OLIS Shutdown. In no particular order:
* DB1/3 optics refurbished.
* IOS:FC3 replaced.
* MWS refurbished.
* MWS microwave generator water interlock fixed.
* IOS:IG1 requires filament. Pending available window to vent.
* IOS:RP1 scroll pump refurbished. Base pressure <<14mTorr.
* IOS:MB vacuum box liners removed.
* IOS overpressure valve installed on DB6. |