OLIS Logbook General
  General  Not logged in ELOG logo
Message ID: 168     Entry time: Thu Oct 22 15:54:24 2009
Author: Damien Gallop 
Type:  
Category:  
Subject: MCIS on with 33S5+ on FC6 overnight cont'd 

The Supernanogan was on overnight quietly. FC6 went in this morning. (Didn't check if fingers were involved.) I  now notice a strange pulsing in our Stripchart. See what you think. Note: field and A/q max scales are orders of magnitude higher than real life.

Later...At around 14:00 I added ferrite cores to the field probe cable at the CANbus interface box. That clipped the high side of the pulses completely. If I play very carefully with the field range, I can get small artifact-looking pulses to appear since I installed them (see 17:16 chart). In the meantime, I found that, contrary to appearances, the field stability program was not really regulating. A remote reboot of the server cured it.

Attachment 1: pulse.png  111 kB  | Show | Show all
Attachment 2: pulse-clip.png  109 kB  Uploaded Thu Oct 22 18:29:26 2009  | Show | Show all
Attachment 3: pulse-tweak.png  125 kB  Uploaded Thu Oct 22 18:30:13 2009  | Show | Show all
ELOG V2.9.2-2455