Message ID: 13     Entry time: Wed Jun 10 11:22:59 2009
Author: Damien Gallop 
Type: Routine 
Category: General 
Subject: Supernanogan development mass scans 

Main event this morning so far: mass scans and post-scan analysis using Gabe's tools. We have a pretty good recipe now. Tight scan to mass 20 in progress. Later... good scan. Geoff and I did a dry run of our presentations for Friday morning. We used the Auditorium computers, etc. After some small teething pains, everything went well, including a cold start-up of the source. The 17O3+ emittance scan was taken quickly. Pretty good outcome anyway. The mass scan shows nice peaks at several known points: 1H+, 2H2, 4He, 7Li, 12C, 14N, 16O, 19F and 20Ne. Gabe's Calibrate tool returns values for the two constants that are pretty close to what is there now: -1 and 0.0419 calculated versus 0 and 0.0418 actual.

Attachment 1: grab.t27875.png  22 kB  Uploaded Wed Jun 10 18:39:03 2009  | Hide | Hide all
grab.t27875.png
Attachment 2: grab.R27900.png  16 kB  Uploaded Wed Jun 10 18:39:19 2009  | Show | Hide all
Attachment 3: grab.H27444.png  31 kB  Uploaded Wed Jun 10 18:39:31 2009  | Show | Hide all
ELOG V2.9.2-2455