OLIS Logbook General
  General  Not logged in ELOG logo
Message ID: 130     Entry time: Mon Aug 17 15:53:54 2009
Author: Damien Gallop 
Type:  
Category:  
Subject: MCIS mass scans & initial emittance scans 

High resolution zero to 25. Save file: masscorr_mcis_090817-highres.txt

Ack! Elog server went down at ~17:00!!!

18O4+ emittance scan at 22.44kV and slits open: 090817_1707ILTEMIT.txt

Note that in preparation for this scan I steered beam in X then in Y onto FC3 and FC6 respectively, looking for all beam through each slit. There was flat-topping evident in both directions for both cups. I would like to repeat the study for well-defined X beam through each slit by scanning with the 1mm slit and noting the drop-off at the tails. For what they're worth, I am appending the stripcharts for each cup. Data files are here:

mcis_strip_steering_fc3_090817.dat

mcis_strip_steering_fc6_090817.dat

Attachment 1: mciscorr_0-25.png  185 kB  | Hide | Hide all
mciscorr_0-25.png
Attachment 2: mciscorr_0-4.png  185 kB  | Show | Hide all
Attachment 3: mciscorr_3-7.png  186 kB  | Show | Hide all
Attachment 4: mciscorr_7-12.png  172 kB  | Show | Hide all
Attachment 5: mciscorr_17-23.png  177 kB  | Show | Hide all
Attachment 6: mcis_rf.png  93 kB  Uploaded Tue Aug 18 15:20:14 2009  | Show | Hide all
Attachment 7: mcis_emit.png  102 kB  Uploaded Tue Aug 18 15:21:41 2009  | Show | Hide all
Attachment 8: mcis_emit_calc.png  126 kB  Uploaded Tue Aug 18 15:21:54 2009  | Show | Hide all
Attachment 9: emit_rig.png  79 kB  Uploaded Tue Aug 18 15:22:07 2009  | Show | Hide all
Attachment 10: steering_fc3.png  67 kB  Uploaded Tue Aug 18 15:34:05 2009  | Show | Hide all
Attachment 11: steering_fc6.png  92 kB  Uploaded Tue Aug 18 15:35:25 2009  | Show | Hide all
ELOG V2.9.2-2455