OLIS Logbook General
  General  Not logged in ELOG logo
Message ID: 121     Entry time: Mon Aug 10 16:43:59 2009
Author: Damien Gallop 
Type:  
Category:  
Subject: MWS 25Mg emittance scan with slits open 

Before proceeding, I nudged in each of the slits, watching for first clipping. With the current emittance (nice), the beam cleared both slits completely; the upstream slit began clipping at about 1.5mm, and the downstream one at about 2mm. Also, FC6 current with slits in/out was: ~30nA / ~57nA. Note: I had to change the scanner parameters a little.

Data file: 090810_1648ILTEMIT.txt

Attachment 1: grab.G23396.png  24 kB  Uploaded Mon Aug 10 17:50:00 2009  | Hide | Hide all
grab.G23396.png
Attachment 2: grab.g23400.png  15 kB  Uploaded Mon Aug 10 17:50:09 2009  | Show | Hide all
Attachment 3: grab.r23404.png  14 kB  Uploaded Mon Aug 10 17:50:19 2009  | Show | Hide all
ELOG V2.9.2-2455