M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 9 of 15  ELOG logo
Entry  Thu Jul 20 00:52:13 2006, Doug Preddy, Quench Recovery, T1 frozen 
T1 will not spin up. Warming up cold box.
Entry  Thu Jul 20 02:59:06 2006, Doug Preddy, Quench Recovery, T1 still not starting 
I will leave the cold box to warm overnight and try it again in the morning.
Entry  Thu Jul 20 08:18:31 2006, Doug Preddy, Quench Recovery, T1 started 
Started flow to turbines. Both started. T1 seems a bit slow. HIC110 at 0.85 
SI110 = 4100.
Cool down started.
Entry  Fri Jul 21 07:39:37 2006, Doug Preddy, Quench Recovery, Cool down continues 
Raised PIC229 to 11Bar. Started to close HIC155. Cooling rates improve.
Entry  Fri Jul 21 14:08:04 2006, Doug Preddy, Quench Recovery, switch to series mode 
Coils temps <10K.Switching to series mode.
Entry  Fri Jul 21 23:36:21 2006, Doug Preddy, Quench Recovery, Final cooldown problems 
80K shield is down to <60K Liquid level not building. Power leads rippling 
and warming. T1 inlet 85K with 10%heater load.
Lowered the set point of TI121 to 17K. Shut off flow to the 80K shield to 
Entry  Fri Jul 28 20:39:40 2006, Doug Preddy, Quench Recovery, Restarting refrigerator 
Solenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold box and turbines. When cold box 
was at 25K I started flow to the solenoid in parallel cold return.
Entry  Sat Jul 29 05:30:28 2006, Doug Preddy, Quench Recovery, switched to series mode 
Temps bottomed out at ~16K. Switching to series mode.
Entry  Sat Jul 29 09:38:06 2006, Doug Preddy, Quench Recovery, power leads cooled 
opened power lead flow about 40 minutes ago. ~12K will let them stablize. 
LHE level >20%
Entry  Tue Aug 8 07:27:49 2006, Doug Preddy, Quench Recovery, Solenoid Quench at about 6:00 this morning 
No indication of problem. Turbines still running, but no flow to solenoid. 
Started flow. Solenoid ~20K.
Entry  Wed Aug 9 07:50:49 2006, Doug Preddy, Quench Recovery, Switched to series mode. 
Cool down continues.
Entry  Thu Aug 10 08:33:44 2006, Doug Preddy, Normal Running, Running low on helium 
Temperture okay, but there are fluctuation in the power leads. PIC229 is 
trying to charge. The buffer tank pressure is just >0 psi. I will lower the 
liquid level to 40% to allow some helium to boil off into the buffer tank.
Entry  Tue Aug 15 08:45:56 2006, Doug Preddy, Quench Recovery, restarted the compressor 
Solenoid tripped off Sunday. Restarted compressor and cool down of cold box.
Entry  Thu Aug 17 08:04:44 2006, Doug Preddy, Quench Recovery, Solenoid tripped at about 3:00 this morning 
restarting system.
Compressor would not start. Idid not see anything that would not let it 
start. Pushed rest button upstairs, and reset contactor. Compressor started.
Entry  Wed Oct 4 07:29:04 2006, Doug Preddy, Start Up, Starting cool down 
Started compressor yesterday with flow through the purifier overnight.
Starting tubines this morning in parallel warm return
Entry  Thu Oct 5 10:26:26 2006, Doug Preddy, Start Up, switched to parallel cold return 
 
Entry  Thu Oct 5 11:51:53 2006, Doug Preddy, Start Up, Adde 2 bottles UHP Helium 
Valved off purifiers, but left LN2 on.
Entry  Thu Oct 5 14:53:54 2006, Doug Preddy, Start Up, Switched to series mode 
 
Entry  Thu Oct 26 08:18:10 2006, Doug Preddy, Quench Recovery, switched to series mode 
Switched to series mode at about 7:45.
Entry  Thu Oct 26 11:45:22 2006, Doug Preddy, Quench Recovery, Solenoid cold 
power leads cold 50% liquid.
Turning on power supply