ID Date Author Operational Mode Subjectdown Text Attachments
  37   Wed Jul 28 23:07:01 2004 Doug PreddyQuench Recoverysolenoid coolingcooling rates for the solenoid are good. In
parallel cold return.
  
  265   Wed Dec 10 15:41:04 2008 Doug PreddyShut Downshuttting down solenoidshuttting down solenoid for winter shutdown.   
  116   Thu Dec 22 15:10:44 2005 Doug PreddyShut Downshut down systemshuttting down the system. isolated cold box
by closing v135. preparing to 
pump out cold box.
  
  91   Fri Mar 11 09:18:30 2005 Doug PreddyMaintenanceservices on systemThe cold box was isolated and pumped out after
it warmed up. It has been 
refilled with helium.
  
  238   Fri Nov 23 08:41:45 2007 Doug PreddyNormal Runningscreen grab of powerlead temps   Capture.jpg 
  264   Wed Nov 5 16:33:52 2008 Doug PreddyNormal Runningsbplc light outOn main menu the sbplc light was black. All
vacuum systems looked good, but there was
no control of any device. Jane reset the
  
  65   Sat Oct 16 17:13:19 2004 Other ss   
  69   Wed Oct 20 09:38:17 2004 Doug PreddyStart Uprestarting systemSolenoid tripped  off Monday morning due to
high compressor pressure. 
Regenerated purifier on Tuesday. Started
  
  158   Tue Aug 15 08:45:56 2006 Doug PreddyQuench Recoveryrestarted the compressorSolenoid tripped off Sunday. Restarted compressor
and cool down of cold box.
  
  103   Tue Jun 14 20:53:19 2005 Doug PreddyQuench Recoveryrestarted refrigerator after power bumpsee bitmap for temps/pressures after presurrizing
coldbox. Preparing to 
start flow to turbines, then will start flow
 june142045.BMP 
  112   Thu Nov 3 06:19:50 2005 Doug PreddyQuench Recoveryrestart systemLast night at about 7:00 there was a large
power bump that shut the 
compressor and water pump off. I came in
  
  188   Fri Apr 20 12:15:12 2007 Doug PreddyStart Upremote PLC switchI could not control the vacuum devices using
EPICS. The buttons could be 
pushed, but nothing would respond. The PLC
  
  216   Wed Sep 12 12:01:02 2007 Doug PreddyMaintenanceregenerated adsorberregenerated adsorber
replaced MV902 "ON" switch (used nitrogen
precooling on switch)
  
  245   Wed Apr 16 15:15:44 2008 Doug PreddyStart Uprefrigerator started Turned on compressor last Tuesday. Ran purifier
until today. Final value 0.15ppm.
Started
cool down yesterday. Switched into parallel
  
  202   Thu Jun 14 00:10:10 2007 Doug PreddyQuench Recoveryrecovering from earlier quenchSolenoid quenched on power lead temperature
last night at about 6:00. I 
restarted flow to the solenoid in parallel
  
  184   Tue Nov 28 10:32:04 2006 Doug PreddyQuench Recoveryrecovering after looong power outageThe power has been out for about 24 hours.
Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum
  
  257   Thu Sep 4 10:02:55 2008 Doug PreddyNormal Runningramping down solenoidSolenoid power supply will be turned off for
Fall shutdown.
  
  209   Sat Jul 28 15:45:17 2007 Doug PreddyQuench Recoveryquench recoveryThe solenoid tripped off this morning due
to pwer lead over temp. The 
compressor was also off. I restarted this
  
  210   Sat Jul 28 22:30:13 2007 Doug PreddyQuench Recoveryquench againrecovered system and energized solenoid at
about 6:00 this evening.
At 9:30 power lead temps quenched solenoid
  
  194   Mon Apr 30 09:30:26 2007 Doug PreddyNormal Runningproblems over weekendThe PLC Limit switch power supply was tripped
off on Saturday. I came in 
and reset it and restarted all the vacuum
  