M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 4 of 15  ELOG logo
ID Date Author Operational Mode Subjectdown
  37   Wed Jul 28 23:07:01 2004 Doug PreddyQuench Recoverysolenoid cooling
cooling rates for the solenoid are good. In parallel cold return.
  265   Wed Dec 10 15:41:04 2008 Doug PreddyShut Downshuttting down solenoid
shuttting down solenoid for winter shutdown.
  116   Thu Dec 22 15:10:44 2005 Doug PreddyShut Downshut down system
shuttting down the system. isolated cold box by closing v135. preparing to 
pump out cold box.
  91   Fri Mar 11 09:18:30 2005 Doug PreddyMaintenanceservices on system
The cold box was isolated and pumped out after it warmed up. It has been 
refilled with helium.
Turbine 2 speed indicator failed at some time during the last run. After 
the turbine inlet vavle was closed the turbine continued to read full 
speed. I am removing the controller to look at it. The main power has been 
turned of for this procedure.
  238   Fri Nov 23 08:41:45 2007 Doug PreddyNormal Runningscreen grab of powerlead temps
Attachment 1: Capture.jpg
Capture.jpg
  264   Wed Nov 5 16:33:52 2008 Doug PreddyNormal Runningsbplc light out
On main menu the sbplc light was black. All vacuum systems looked good, but there was no control of any device. Jane reset the IOC. It is hosted by sundance.
  65   Sat Oct 16 17:13:19 2004 Other s
s
  69   Wed Oct 20 09:38:17 2004 Doug PreddyStart Uprestarting system
Solenoid tripped  off Monday morning due to high compressor pressure. 
Regenerated purifier on Tuesday. Started compressor at 9:00 this morning.
  158   Tue Aug 15 08:45:56 2006 Doug PreddyQuench Recoveryrestarted the compressor
Solenoid tripped off Sunday. Restarted compressor and cool down of cold box.
  103   Tue Jun 14 20:53:19 2005 Doug PreddyQuench Recoveryrestarted refrigerator after power bump
see bitmap for temps/pressures after presurrizing coldbox. Preparing to 
start flow to turbines, then will start flow to solenoid in parallel warm 
return.
Attachment 1: june142045.BMP
  112   Thu Nov 3 06:19:50 2005 Doug PreddyQuench Recoveryrestart system
Last night at about 7:00 there was a large power bump that shut the 
compressor and water pump off. I came in this morning and am restarting the 
system.
  188   Fri Apr 20 12:15:12 2007 Doug PreddyStart Upremote PLC switch
I could not control the vacuum devices using EPICS. The buttons could be 
pushed, but nothing would respond. The PLC was in local mode. Thumbwheel 
to 100, push On/Open button to put it into remote mode.
  216   Wed Sep 12 12:01:02 2007 Doug PreddyMaintenanceregenerated adsorber
regenerated adsorber
replaced MV902 "ON" switch (used nitrogen precooling on switch)
  245   Wed Apr 16 15:15:44 2008 Doug PreddyStart Uprefrigerator started
Turned on compressor last Tuesday. Ran purifier until today. Final value 0.15ppm. Started cool down yesterday. Switched into parallel cold return this morning, but PIC 229 would not allow the compressor to charge. The compressor tripped due to bad solenoid vacuum (solenoid was warming up). The controller now seems to be working okay. Restarted in parallel warm mode at ~12:00
  202   Thu Jun 14 00:10:10 2007 Doug PreddyQuench Recoveryrecovering from earlier quench
Solenoid quenched on power lead temperature last night at about 6:00. I 
restarted flow to the solenoid in parallel cold return. Switched to series 
mode and am building liquid. 
  184   Tue Nov 28 10:32:04 2006 Doug PreddyQuench Recoveryrecovering after looong power outage
The power has been out for about 24 hours. Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum system. Started compressor and 
am flowing warm helium through the soleniod to warm it up while the vacuum 
recovers.
  257   Thu Sep 4 10:02:55 2008 Doug PreddyNormal Runningramping down solenoid
Solenoid power supply will be turned off for Fall shutdown.
  209   Sat Jul 28 15:45:17 2007 Doug PreddyQuench Recoveryquench recovery
The solenoid tripped off this morning due to pwer lead over temp. The 
compressor was also off. I restarted this morning at about 8:15 in 
parallel cold return. Switching to series mode at 3:40.
  210   Sat Jul 28 22:30:13 2007 Doug PreddyQuench Recoveryquench again
recovered system and energized solenoid at about 6:00 this evening.
At 9:30 power lead temps quenched solenoid again. MV52 stuck open during 
start-up.
  194   Mon Apr 30 09:30:26 2007 Doug PreddyNormal Runningproblems over weekend
The PLC Limit switch power supply was tripped off on Saturday. I came in 
and reset it and restarted all the vacuum systems that had tripped off. 
The power leads were found to be warming up. This morning they were at ~ 
45K. I stopped flow to the power leads and will allow them to warm up 
above 80K and then will recool them.