Entry  Wed Jul 28 23:07:01 2004, Doug Preddy, Quench Recovery, solenoid cooling 
cooling rates for the solenoid are good. In parallel cold return.
Entry  Wed Dec 10 15:41:04 2008, Doug Preddy, Shut Down, shuttting down solenoid 
shuttting down solenoid for winter shutdown.
Entry  Thu Dec 22 15:10:44 2005, Doug Preddy, Shut Down, shut down system 
shuttting down the system. isolated cold box by closing v135. preparing to 
pump out cold box.
Entry  Fri Mar 11 09:18:30 2005, Doug Preddy, Maintenance, services on system 
The cold box was isolated and pumped out after it warmed up. It has been 
refilled with helium.
Turbine 2 speed indicator failed at some time during the last run. After 
Entry  Fri Nov 23 08:41:45 2007, Doug Preddy, Normal Running, screen grab of powerlead temps Capture.jpg
 
Entry  Wed Nov 5 16:33:52 2008, Doug Preddy, Normal Running, sbplc light out 
On main menu the sbplc light was black. All vacuum systems looked good, but there was no control of any device. Jane reset the IOC. It is hosted by sundance.
Entry  Sat Oct 16 17:13:19 2004, Other, , s 
s
Entry  Wed Oct 20 09:38:17 2004, Doug Preddy, Start Up, restarting system 
Solenoid tripped  off Monday morning due to high compressor pressure. 
Regenerated purifier on Tuesday. Started compressor at 9:00 this morning.
Entry  Tue Aug 15 08:45:56 2006, Doug Preddy, Quench Recovery, restarted the compressor 
Solenoid tripped off Sunday. Restarted compressor and cool down of cold box.
Entry  Tue Jun 14 20:53:19 2005, Doug Preddy, Quench Recovery, restarted refrigerator after power bump june142045.BMP
see bitmap for temps/pressures after presurrizing coldbox. Preparing to 
start flow to turbines, then will start flow to solenoid in parallel warm 
return.
Entry  Thu Nov 3 06:19:50 2005, Doug Preddy, Quench Recovery, restart system 
Last night at about 7:00 there was a large power bump that shut the 
compressor and water pump off. I came in this morning and am restarting the 
system.
Entry  Fri Apr 20 12:15:12 2007, Doug Preddy, Start Up, remote PLC switch 
I could not control the vacuum devices using EPICS. The buttons could be 
pushed, but nothing would respond. The PLC was in local mode. Thumbwheel 
to 100, push On/Open button to put it into remote mode.
Entry  Wed Sep 12 12:01:02 2007, Doug Preddy, Maintenance, regenerated adsorber 
regenerated adsorber
replaced MV902 "ON" switch (used nitrogen precooling on switch)
Entry  Wed Apr 16 15:15:44 2008, Doug Preddy, Start Up, refrigerator started  
Turned on compressor last Tuesday. Ran purifier until today. Final value 0.15ppm.
Started cool down yesterday. Switched into parallel cold return this
morning, but PIC 229 would not allow the compressor to charge. The compressor tripped due to bad solenoid vacuum (solenoid was warming up). The controller
now seems to be working okay. 
Restarted in parallel warm mode at ~12:00
Entry  Thu Jun 14 00:10:10 2007, Doug Preddy, Quench Recovery, recovering from earlier quench 
Solenoid quenched on power lead temperature last night at about 6:00. I 
restarted flow to the solenoid in parallel cold return. Switched to series 
mode and am building liquid. 
Entry  Tue Nov 28 10:32:04 2006, Doug Preddy, Quench Recovery, recovering after looong power outage 
The power has been out for about 24 hours. Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum system. Started compressor and 
am flowing warm helium through the soleniod to warm it up while the vacuum 
Entry  Thu Sep 4 10:02:55 2008, Doug Preddy, Normal Running, ramping down solenoid 
Solenoid power supply will be turned off for Fall shutdown.
Entry  Sat Jul 28 15:45:17 2007, Doug Preddy, Quench Recovery, quench recovery 
The solenoid tripped off this morning due to pwer lead over temp. The 
compressor was also off. I restarted this morning at about 8:15 in 
parallel cold return. Switching to series mode at 3:40.
Entry  Sat Jul 28 22:30:13 2007, Doug Preddy, Quench Recovery, quench again 
recovered system and energized solenoid at about 6:00 this evening.
At 9:30 power lead temps quenched solenoid again. MV52 stuck open during 
start-up.
Entry  Mon Apr 30 09:30:26 2007, Doug Preddy, Normal Running, problems over weekend 
The PLC Limit switch power supply was tripped off on Saturday. I came in 
and reset it and restarted all the vacuum systems that had tripped off. 
The power leads were found to be warming up. This morning they were at ~ 