266
Wed Dec 10 15:44:58 2008
Doug Preddy
Maintenance
Repairs to solenoid power supply
Klaus and company pulled the ramp rate card
out of the solenoid power supply. Operational
checks were done and the rate switch was
265
Wed Dec 10 15:41:04 2008
Doug Preddy
Shut Down
shuttting down solenoid
shuttting down solenoid for winter shutdown.
264
Wed Nov 5 16:33:52 2008
Doug Preddy
Normal Running
sbplc light out
On main menu the sbplc light was black. All
vacuum systems looked good, but there was
no control of any device. Jane reset the
263
Sun Oct 5 18:54:27 2008
Doug Preddy
Quench Recovery
Solenoid Quench ~16:30
Restarted compressor and flow to coldbox by
18:30
262
Sun Oct 5 11:12:48 2008
Doug Preddy
Normal Running
M9B:V1 interlocks and a IOC crash
M9A:V1 was removed this shutdown. It's "in"
limit is required to allow M9B:V1 to be opened.
I have bypassed the interlocks on M9B:V1
261
Fri Sep 26 09:50:18 2008
Doug Preddy
Quench Recovery
started flow in parallel cold return mode
260
Fri Sep 26 09:45:18 2008
Doug Preddy
Quench Recovery
Solenoid quenched at 7:50
Compressor on. Turbines started. Vacuum system
recovering.
259
Fri Sep 26 09:44:25 2008
Doug Preddy
Maintenance
Water flow interlock restored
yesterday
258
Tue Sep 9 15:04:23 2008
Doug Preddy
Maintenance
Jumpered interlock for diff pump water flow
Jumpered interlock for diff pump water flow
in Proteus box.
257
Thu Sep 4 10:02:55 2008
Doug Preddy
Normal Running
ramping down solenoid
Solenoid power supply will be turned off for
Fall shutdown.
256
Sun Aug 24 06:45:14 2008
Doug Preddy
Quench Recovery
switch to series mode
255
Sat Aug 23 17:02:38 2008
Doug Preddy
Quench Recovery
Solenoid quenched
The solenoid quenched this afternoon at ~12:30.
The compressor shut down. Fault lights were
cold box trip.
Restarted the compressor
254
Sat Jul 26 01:29:21 2008
Doug Preddy
Quench Recovery
power leads
ops will monitor liquid level. When it's at
60% they will come down and open the power
lead valve. When the temperature ripples
253
Sat Jul 26 01:06:04 2008
Doug Preddy
Quench Recovery
Refrigerator in series mode
Switched to series mode. Solenoid ~8K, starting
to build liquid.
252
Fri Jul 25 21:48:15 2008
Doug Preddy
Quench Recovery
another quench
System was stable and solenoid ramped up at
4:00 today. At about 7:30 it tripped off
again. I came in and restarted the compressor
251
Thu Jul 24 16:28:50 2008
Doug Preddy
Quench Recovery
power bump
There was a power bump at ~15:00 today. All
mseon hall cooling water went off. Compressor
tripped. Main magnet went off.
Restarted
250
Tue Jul 15 08:47:13 2008
Doug Preddy
Normal Running
Screen grab of normal running
Solenoid was recovered on Sunday July 13.
The power leads started to ripple slightly
this morning, so I cycled HIC142. No real
249
Sun Jul 13 20:17:24 2008
Doug Preddy
Other
Helium refrigerator problems
In early June the helium compressor was running
with the solenoid cold. A vavlve broke off
at the suction side of the compressor. The
248
Mon Apr 28 09:03:22 2008
Doug Preddy
Normal Running
power leads warmed during weekend
The power lead temperatures suddenly rose
on Sunday morning. No rippling occured before
the woarm up.
I shut flow off this morning
247
Thu Apr 17 13:52:25 2008
Doug Preddy
Start Up
switched to series mode at 9:30 this morning
switched to series mode at 9:30 this morning.
Added two bottles of helium gas.
Started
flow to power leads.