ID Date Author Operational Mode Subjectdown Text Attachments
  263   Sun Oct 5 18:54:27 2008 Doug PreddyQuench RecoverySolenoid Quench ~16:30Restarted compressor and flow to coldbox by
18:30
  
  115   Mon Dec 12 09:24:29 2005 Doug PreddyQuench RecoverySolenoid Quench on power lead tempThe solenoid tripped on power lead temps at
about 9:15. I am recovering it 
from it's all around mode into parallel cold
  
  151   Tue Aug 8 07:27:49 2006 Doug PreddyQuench RecoverySolenoid Quench at about 6:00 this morningNo indication of problem. Turbines still running,
but no flow to solenoid. 
Started flow. Solenoid ~20K.
  
  227   Tue Oct 9 12:58:22 2007 Doug PreddyQuench RecoverySolenoid Quench 2Refrigeration sytem shut off at ~ 12:30 today.
No apparent reason. Only 
light on was a cold box fault. 
  
  191   Fri Apr 27 10:10:07 2007 Doug PreddyNormal RunningSolenoid ColdSolenoid cold, power leads cold. Alarm handler
started.
  
  15   Mon Apr 26 07:48:46 2004 Doug PreddyQuench RecoverySeries modeSwitching over to series mode.   
  250   Tue Jul 15 08:47:13 2008 Doug PreddyNormal RunningScreen grab of normal runningSolenoid was recovered on Sunday July 13.
The power leads started to ripple slightly
this morning, so I cycled HIC142. No real
 july_15_08_0841.jpgjuly_15_08_0846.jpg 
  93   Thu May 19 15:05:31 2005 Doug PreddyMaintenanceSTA 120 replacedThe turbine 2 speed indicator control box
STA 120 has been replaced with a 
newer style Jaquet controller. This is similar
 M9 Turbine 2.doc 
  155   Thu Aug 10 08:33:44 2006 Doug PreddyNormal RunningRunning low on heliumTemperture okay, but there are fluctuation
in the power leads. PIC229 is 
trying to charge. The buffer tank pressure
  
  148   Fri Jul 28 20:39:40 2006 Doug PreddyQuench RecoveryRestarting refrigeratorSolenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold
box and turbines. When cold box 
  
  193   Fri Apr 27 10:32:07 2007 Doug PreddyQuench RecoveryReset quench relay logic boxThe quench relay logic controller was not
functioning correctly. It showed 
a "Trip" light and no other lights. I cycled
  
  266   Wed Dec 10 15:44:58 2008 Doug PreddyMaintenanceRepairs to solenoid power supplyKlaus and company pulled the ramp rate card
out of the solenoid power supply. Operational
checks were done and the rate switch was
  
  31   Tue Jun 29 14:59:46 2004 Willy AnderssonNormal RunningRemoved jumper on vacuum controller.Tuesday 14:55

Removed jumper on back of PIC 901. This jumper
  
  92   Mon Apr 18 09:52:11 2005 Doug PreddyMaintenanceRelays replacedThe following relays have been replaced in
the Sulzer control panel:
K341, K379, K132, K324A, K314, K125, K127,
  
  253   Sat Jul 26 01:06:04 2008 Doug PreddyQuench RecoveryRefrigerator in series modeSwitched to series mode. Solenoid ~8K, starting
to build liquid. 
  
  122   Fri May 26 14:34:24 2006 Doug PreddyStart UpReady to runSolenoid is cold, liquid level is 35%, power
leads cold and seem stable.
Added one bottle helium to buffer tank. Pressure
  
  163   Wed Oct 4 18:10:18 2006 Willy AnderssonStart UpQuick check on systemSystem in good // warm return mode, temps
coming down nicely. BCP2 cryopump 
does not look too healthy, temperature is
  
  49   Mon Sep 13 10:55:25 2004 Doug PreddyShut DownQuench and shutdownThe solenoid quenched on powerlead temps this
morning at 10:00. It is 
warming up and the power leads are >200K.Isolating
  
  108   Tue Jul 19 18:09:56 2005 Willy AnderssonQuench RecoveryQuench IIMachine trip just as I was exiting the TRIUMF
gate. QRL said cold box fault 
as did red lite on compressor. At cold box
  
  198   Wed May 30 11:27:32 2007 Doug PreddyQuench RecoveryProblems recoveringVarious start up problems :
MV51 stuck open. Finally got it closed by
hitting the valve with a hammer.
  