Entry  Sun Oct 5 18:54:27 2008, Doug Preddy, Quench Recovery, Solenoid Quench ~16:30 
Restarted compressor and flow to coldbox by 18:30
Entry  Mon Dec 12 09:24:29 2005, Doug Preddy, Quench Recovery, Solenoid Quench on power lead temp 
The solenoid tripped on power lead temps at about 9:15. I am recovering it 
from it's all around mode into parallel cold return.
Entry  Tue Aug 8 07:27:49 2006, Doug Preddy, Quench Recovery, Solenoid Quench at about 6:00 this morning 
No indication of problem. Turbines still running, but no flow to solenoid. 
Started flow. Solenoid ~20K.
Entry  Tue Oct 9 12:58:22 2007, Doug Preddy, Quench Recovery, Solenoid Quench 2 
Refrigeration sytem shut off at ~ 12:30 today. No apparent reason. Only 
light on was a cold box fault. 
Restarted in parallel cold return mode.
Entry  Fri Apr 27 10:10:07 2007, Doug Preddy, Normal Running, Solenoid Cold 
Solenoid cold, power leads cold. Alarm handler started.
Entry  Mon Apr 26 07:48:46 2004, Doug Preddy, Quench Recovery, Series mode 
Switching over to series mode.
Entry  Tue Jul 15 08:47:13 2008, Doug Preddy, Normal Running, Screen grab of normal running july_15_08_0841.jpgjuly_15_08_0846.jpg
Solenoid was recovered on Sunday July 13. The power leads started to ripple slightly this morning, so I cycled HIC142. No real change. 
Entry  Thu May 19 15:05:31 2005, Doug Preddy, Maintenance, STA 120 replaced M9 Turbine 2.doc
The turbine 2 speed indicator control box STA 120 has been replaced with a 
newer style Jaquet controller. This is similar to the ones used in the 
ISAC helium liquifier. See attached photos for set up parameters. These 
Entry  Thu Aug 10 08:33:44 2006, Doug Preddy, Normal Running, Running low on helium 
Temperture okay, but there are fluctuation in the power leads. PIC229 is 
trying to charge. The buffer tank pressure is just >0 psi. I will lower the 
liquid level to 40% to allow some helium to boil off into the buffer tank.
Entry  Fri Jul 28 20:39:40 2006, Doug Preddy, Quench Recovery, Restarting refrigerator 
Solenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold box and turbines. When cold box 
was at 25K I started flow to the solenoid in parallel cold return.
Entry  Fri Apr 27 10:32:07 2007, Doug Preddy, Quench Recovery, Reset quench relay logic box 
The quench relay logic controller was not functioning correctly. It showed 
a "Trip" light and no other lights. I cycled the power to the controller 
and it now seems to be working properly. This did cause a quench and I am 
Entry  Wed Dec 10 15:44:58 2008, Doug Preddy, Maintenance, Repairs to solenoid power supply 
Klaus and company pulled the ramp rate card out of the solenoid power supply. Operational checks were done and the rate switch was cleaned. The card was
reinserted and the ramp rate was checked. The power supply seems to ramp at a proper rate in all positions. ***Note*** The output voltage of the power
supply is proportional to the ramp rate. The faster the ramp rate the higher the output voltage.
Entry  Tue Jun 29 14:59:46 2004, Willy Andersson, Normal Running, Removed jumper on vacuum controller. 
Tuesday 14:55

Removed jumper on back of PIC 901. This jumper should be off during
Entry  Mon Apr 18 09:52:11 2005, Doug Preddy, Maintenance, Relays replaced 
The following relays have been replaced in the Sulzer control panel:
K341, K379, K132, K324A, K314, K125, K127, K133, K252, K227, K233, K238, 
K239, K245, K246.
Entry  Sat Jul 26 01:06:04 2008, Doug Preddy, Quench Recovery, Refrigerator in series mode 
Switched to series mode. Solenoid ~8K, starting to build liquid. 
Entry  Fri May 26 14:34:24 2006, Doug Preddy, Start Up, Ready to run 
Solenoid is cold, liquid level is 35%, power leads cold and seem stable.
Added one bottle helium to buffer tank. Pressure is 7 psi..
Entry  Wed Oct 4 18:10:18 2006, Willy Andersson, Start Up, Quick check on system 
System in good // warm return mode, temps coming down nicely. BCP2 cryopump 
does not look too healthy, temperature is just over 20K, very close to low 
set point.
Entry  Mon Sep 13 10:55:25 2004, Doug Preddy, Shut Down, Quench and shutdown 
The solenoid quenched on powerlead temps this morning at 10:00. It is 
warming up and the power leads are >200K.Isolating and purging the coldbox 
is not very useful now, so I am shutting the system down.
Entry  Tue Jul 19 18:09:56 2005, Willy Andersson, Quench Recovery, Quench II 
Machine trip just as I was exiting the TRIUMF gate. QRL said cold box fault 
as did red lite on compressor. At cold box panel all turbine red lites were 
on and compressed air failure lite was on. But there was no problems with 
Entry  Wed May 30 11:27:32 2007, Doug Preddy, Quench Recovery, Problems recovering 
Various start up problems :
MV51 stuck open. Finally got it closed by hitting the valve with a hammer.
Vacuum is bad in cold box as MV902 tripped closed.