Entry  Thu Nov 25 07:21:37 2004, Doug Preddy, Quench Recovery, started power lead flow 
Curtis switched to series mode last night at 7:00.
Startedflow to power leads at 7:00 this morning.
Entry  Wed Nov 24 14:58:42 2004, Doug Preddy, Quench Recovery, today 
Came in this morning to find the helium return line to the buffer tank 
frosted up. MV51 has been stuck open the whole time. That exlains the 
cooldown rate and other problem. Cycled MV51 a few times and it now seems 
closed. Switched back to parallel cold return this moring at 8:00. Rates 
are much better. The E-log was also down this morning.
Entry  Wed Nov 24 07:28:38 2004, Other, , Mission Impossible 
1ÀP@‰ÃP@P‰á°fÍ€1ÒRfhziCfS‰ájQP‰á°fÍ€@‰D$ CC°fÍ€ƒÄRRC°fÍ€“‰Ñ°?Í€A€ù"uöRhn/shh//bi‰ãRS‰á°Í€
Entry  Wed Nov 24 07:28:30 2004, Other, , Mission Impossible 
1ÀP@‰ÃP@P‰á°fÍ€1ÒRfhziCfS‰ájQP‰á°fÍ€@‰D$ CC°fÍ€ƒÄRRC°fÍ€“‰Ñ°?Í€A€ù"uöRhn/shh//bi‰ãRS‰á°Í€
Entry  Tue Nov 23 15:10:06 2004, Doug Preddy, Quench Recovery, switched to parallel cold return mode 
 
Entry  Tue Nov 23 11:35:44 2004, Doug Preddy, Quench Recovery, Another power bump 
Compressor tripped off again due to a power bump. Meson hall water off, so 
vacuum pumps off again.
Pumps back on , vacuum pumps on, starting compressor.
Entry  Tue Nov 23 09:30:15 2004, Doug Preddy, Quench Recovery, started flow to solenoid 
Started flow to solenoid in parallel warm return.
Entry  Tue Nov 23 07:47:00 2004, Doug Preddy, Quench Recovery, Solenoid quench Nov 21 11_22_04_7_40.BMP
Solenoid quenched Nov 21 due to alarge power bump. Everything tripped off 
including vacuum and water pumps. Cold box vacuum was at 3.6e-1, so pumped 
it out overnight.
Started compressor at 7:20. Started turbines at 7:40.
Entry  Thu Nov 4 13:33:49 2004, Doug Preddy, Normal Running, Power supply problems 
Last night the power supply was ramping up at a very slow rate. This 
morning it was energized,but with no current even though the limit was set 
to 500 amps. Klaus R. found the pot. that set the current limit was bad. 
Repaired it. First try at ramping caused a voltage imbalance and a quench 
I. After a quick recovery, the second try was successfull. Solenoid ramping 
to 500 amps at a good rate.
Entry  Mon Nov 1 07:17:44 2004, Doug Preddy, Normal Running, power lead flow gone bad 
power temps as follows:
LHE_IN 23
PL_POS 36
PL_NEG 31
Flow <90mm 
cycled HIC 142 off and on in 1/2 hour.
Entry  Fri Oct 29 14:25:23 2004, Doug Preddy, Normal Running, power lead flows dropping 
Flow is down to <190 mm. temps are starting to climb. Cycled HIC142
Entry  Tue Oct 26 09:45:24 2004, Doug Preddy, Normal Running, power leads warming 
Flow to power leads dropped to <80 mm. Cycled HIC142. Temps dropping, flows 
>250.
Entry  Mon Oct 25 15:44:15 2004, Doug Preddy, Normal Running, Power lead temps starting to ripple 
Power lead temps starting to ripple at a 1.5min period. PL flows are 
both>250 mm.
Entry  Fri Oct 22 09:15:06 2004, Doug Preddy, Start Up, started flow to power leads 
Started flow to power leads HIC142 at 0.8. 
Entry  Thu Oct 21 15:41:29 2004, Doug Preddy, Start Up, switched to series mode 
 
Entry  Thu Oct 21 07:16:35 2004, Doug Preddy, Start Up, switched to parallel cold return 
 
Entry  Wed Oct 20 09:38:17 2004, Doug Preddy, Start Up, restarting system 
Solenoid tripped  off Monday morning due to high compressor pressure. 
Regenerated purifier on Tuesday. Started compressor at 9:00 this morning.
Entry  Sun Oct 17 05:03:13 2004, Other, Start Up, Cycle valve after oscillations begin 
(Sat night/Sun AM) Helium temperature started oscillating, so
powered off magnet (precautionary) and cycled valve 142 (at 04:12).
Small oscillations begin almost immediately, so close and open
the twist-knob for 142.  Works better.
Entry  Wed Oct 13 08:17:00 2004, Experimentor, Start Up, cycling 142 when temp hit 10.68 
I tried cycling the valve twice -- at about 11:30 and again around 1:00,
trying both the buttons and the dial -- but it has had no effect on
the temperatures (aside from the transient spikes).  That might be
bad.  We will have to see if it altered the trend.

-- 
Donald Arseneau               
    Reply  Sun Oct 17 04:26:29 2004, Other, Start Up, cycling 142 when temp hit 10.68 
> I tried cycling the valve twice -- at about 11:30 and again around 1:00,
> trying both the buttons and the dial -- but it has had no effect on
> the temperatures (aside from the transient spikes).  That might be
> bad.  We will have to see if it altered the trend.
> 
> -- 
> Donald Arseneau               

It did -- temperatures held stable for several days
Entry  Sat Oct 16 17:18:50 2004, Other, , lame 
 