ID Date Author Operational Modedown Subject Text Attachments
  252   Fri Jul 25 21:48:15 2008 Doug PreddyQuench Recoveryanother quenchSystem was stable and solenoid ramped up at
4:00 today. At about 7:30 it tripped off
again. I came in and restarted the compressor
  
  253   Sat Jul 26 01:06:04 2008 Doug PreddyQuench RecoveryRefrigerator in series modeSwitched to series mode. Solenoid ~8K, starting
to build liquid. 
  
  254   Sat Jul 26 01:29:21 2008 Doug PreddyQuench Recoverypower leadsops will monitor liquid level. When it's at
60% they will come down and open the power
lead valve. When the temperature ripples
  
  255   Sat Aug 23 17:02:38 2008 Doug PreddyQuench RecoverySolenoid quenchedThe solenoid quenched this afternoon at ~12:30.
The compressor shut down. Fault lights were
cold box trip.
Restarted the compressor
  
  256   Sun Aug 24 06:45:14 2008 Doug PreddyQuench Recoveryswitch to series mode    
  260   Fri Sep 26 09:45:18 2008 Doug PreddyQuench RecoverySolenoid quenched at 7:50Compressor on. Turbines started. Vacuum system
recovering.
  
  261   Fri Sep 26 09:50:18 2008 Doug PreddyQuench Recoverystarted flow in parallel cold return mode    
  263   Sun Oct 5 18:54:27 2008 Doug PreddyQuench RecoverySolenoid Quench ~16:30Restarted compressor and flow to coldbox by
18:30
  
  267   Wed May 27 08:06:00 2009 Doug PreddyQuench RecoverySolenoid quenched at 5:00The solenoid quenched at 5:00 this morning
due to power lead temps. The temps started
oscillating about 10 minutes before the trip.
  
  268   Wed May 27 13:37:22 2009 Doug PreddyQuench Recoveryswitch to series modeSwitched to series mode at ~11:00. Built
liquid to 55% by 13:30. Started flow to the
power leads.
  
  269   Wed May 27 15:01:34 2009 Doug PreddyQuench Recoverypower leadsPower leads are cold. There is still a
small ripple. The power supply can be ramped
up as soon as the ripples die out.
  
  270   Fri Jul 10 00:34:03 2009 Doug PreddyQuench Recoverysolewnoid quench due to power bumpSolenoid quench due to power bump caused
by an airplane crash in Richmond at ~ 22:00.
Compressor off. Water pump off. Restarted
  
  271   Fri Jul 10 01:10:59 2009 Doug PreddyQuench Recoveryswitch to parallel warm return    
  272   Tue Jul 14 15:28:36 2009 Doug PreddyQuench RecoverySolenoid ramped upThe solenoid is cold and I have started
to ramp up the power supply. Some slight
ripples in the power leads, but they are
  
  277   Fri Apr 9 07:34:55 2010 Doug PreddyQuench Recoveryswitched to parallel cold returnSwitched to parallel cold return at about
7:30 this morning
 
  
  279   Mon May 3 07:49:54 2010 Doug PreddyQuench RecoverySolenoid quench SaturdaySolenoid quenched at ~13:30 Saturday. Kevin
came in later in the day to recover it. System
was in all around mode, no flow to solenoid.
  
  280   Thu May 13 07:39:55 2010 Doug PreddyQuench Recoverysolenoid quench last nightSolenoid quenched on power lead temps last
evening. I am recovering it in parallel cold
return mode. C12~60K C1~50K.
  
  281   Thu May 13 14:37:13 2010 Doug PreddyQuench Recoveryswitching to series mode.    
  283   Mon Jun 7 09:24:47 2010 Doug PreddyQuench Recoverystarted flow to power leads    
  99   Thu Jun 9 06:53:40 2005 ExperimentorOtherTrip at 6:45 Today. Solenoid quenched at 6:45.  All temperature
monitors were flat until trip.  
Hearing compressor stop was first warning.
  