Subject
*
:
> Got page very early this morning. Couldn't come in til 08:30. System was > in idle running mode and someone had reset the QRL panel so my guess is > that the solenoid tripped on powerlead overtemp. HIC142 was not fully open > and I suspect this is why powerleads heated up. Now in // cold return mode. > Solenoid temps down to 25K and slowly going down. Have HIC142 slightly open > to keep powerleads cool. Strongly recommend that when running HIC142 should > be fully open to try to reduce the probability of powerlead warmup.