> Came in this morning to find the helium return line to the buffer tank > frosted up. MV51 has been stuck open the whole time. That exlains the > cooldown rate and other problem. Cycled MV51 a few times and it now seems > closed. Switched back to parallel cold return this moring at 8:00. Rates > are much better. The E-log was also down this morning.